AEG FT 4919: English

English: AEG FT 4919

68

English

FT 4919 AEG Ear Thermometer (TH520B)

Technical

Temperature-Measurement range:

Data

Measurements in the “ear area” : 34~42.2°C (93.2~108°F),

Scan- operation: -22~80°C (-7.6~176°F).

Operation temperature range: 10~40°C (50~104°F), relative

humidity ≤85%

Storage temperature range: -20~50°C (-4~122°F), relative

humidity ≤85%.

Transport temperature should be lower than 70°C, relative

humidity ≤95%.

Atmospheric pressure: 800~1013 hPa

In accordance to ASTM E1965-98, EN12470-5:2003

Clinical Thermometers Part 5: Performance of Infra-red ear

thermometers (with maximum device), IEC/EN60601-1-2

(EMC), IEC/EN60601-1 Safety Guidlines.

Deviations

With measurements in the ear: +/-0.2°C (0.4°F) within the

range of 35.5~42°C (95.9~107.6°F),

+/-0.3°C (0.5°F) out of this range.

In Scan-mode +/-0.3°C (0.5°F) within the range of

22~42.2°C (71.6~108°F),

+/-4% or +/-2°C (4°F) out of this range—

the higher value is decisive.

Fever Alarm saving locations for ear operation (display of

date/time/operation mode for each measurement).

Time display, °C / °F toggle function.

Battery, 1 Lithium cell (CR2032 1 pc.).

Battery life: around 3,000 readings. Enable normal use: 1 year

Expected Service Life: 4 years

Measurement in the ear, the device will calculate the

corresponding value which would be derived from an oral

measurement.

Battery low warning.

Automatic switch-off of measurement activation.

Enclosure Rating: IP22

Dimensions: 146.5 x 38.2 x 54.7 mm

Weight: 55.5 grams including battery

The device should not submerge into any liquids and

expose it to direct moisture.

This symbol indicates important for persons or the

device!