Beurer BM 45: 10. Specifications 9. Rectifying faults

10. Specifications 9. Rectifying faults: Beurer BM 45

Do not use detergents or solvents.

10. Specifications

On no account must you immerse the computer in water,

Model no. BM 45

otherwise liquid can enter it and cause demage.

Type M1502

9. Rectifying faults

Measurement

Oscillometric, non-invasive blood pres-

In case of faults, the _ message appears in the display.

method

sure measurement on the upper arm

Error messages may appear if:

Measurement range Cuff pressure 0 300 mmHg,

systolic or diastolic pressure could not be measured ( or

systolic 60 – 260 mmHg,

appears on the display),

diastolic 40 –199 mmHg,

systolic or diastolic pressure was outside the measurement

Pulse 30 –180 beats/minute

range ( or Lo appears on the display),

Display accuracy Systolic ± 3 mmHg, diastolic ± 3 mmHg,

the cuff is fastened too tightly or loosely ( or appears

pulse ± 5 % of the value shown

on the display),

the pump pressure is higher than 300 mmHg ( appears

Measurement

Max. permissible standard deviation

on the display),

inaccuracy

according to clinical testing:

pumping up takes longer than 160 seconds ( appears on

systolic 8 mmHg /diastolic 8 mmHg

the display),

Memory 2 x 60 memory spaces

there is a system or device error ( , , or ap-

Dimensions L 165 mm x W 107 mm x H 50 mm

pears on the display),

Weight Approx. 284 g (without batteries)

the batteries are almost empty ,

the data could not be sent to the PC ( appears in the

Cuff size 22 to 36 cm

display).

Permissible operating

+10 °C to + 40 °C, ≤ 90 % relative air

conditions

humidity (non-condensing)

In such cases, repeat the measurement. Ensure that the cuff

line is correctly attached and that you do not move or speak. If

Permissible storage

- 20 °C to + 55 °C, ≤ 90 % relative air hu-

necessary, reinsert or replace the batteries.

and transport condi-

midity, 800 –1050 hPa ambient pressure

tions

Power supply

4 x 1,5 V

AA batteries

Battery life For approx. 250 measurements, de-

pending on the blood pressure level

and/or pump pressure

26