Error message Cause Solution
12. Technical data
Take out the battery, wait
for one minute and then
The device is not work-
Temperature range 34.0 – 42.2 °C (93.2 – 108.0 °F)
reinsert the battery. If the
ing properly.
message is shown again,
Measurement accuracy
In the measurement range of 35.0 to
contact your retailer.
42.2 °C (95.0 to 108.0 °F): +/-0.2 °C (0.4 °F),
in other indicating ranges: +/-0.3 °C (0.5 °F)
The device cannot be
Operating temperature
10 – 40 °C (50 – 104 °F), humidity ≤ 95%;
switched on and will
Insert a new battery.
range
700 – 1060hPa
not switch to standby.
Storage temperature range -25 – 55 °C (-13 – 131 °F), humidity ≤95%;
700 – 1060hPa
Check whether the protec-
The measured tem-
tive cover for the measur-
Memory 9 memory spaces
perature is over 42.2 °C
ing probe is damaged and
Display resolution 0.1
(108.0 °F).
take a new temperature
measurement.
Battery 1 x 3V CR2032 battery
Check whether the protec-
Weight 72 g including battery
The measured tem-
tive cover for the measur-
perature is under 34 °C
ing probe is clean and take
Dimensions 153 x 42 x 35 mm
(93.2 °F).
a new temperature meas-
Protective cap 6 pcs
urement.
Automatic switch-o 60 seconds
Battery low; further
measurements are not
Replace the battery.
Battery operating life 3,000 measurements in succession or one
possible.
year with one to two measurements per
day, including standby mode.
Depleted battery: if
Safety classi cation
Connected medical device of type
the battery symbol is
BF
shown constantly, no
Replace the battery.
Clinical reproducibility 0.08 °C (< 1 year old), 0.10 °C (1 – 5 years
more measurements
old), 0.07 °C (>5 years old)
are possible.
IP classi cation IP 22, protection against harmful ingress of
If you cannot nd a solution to your problem here, contact our Cus-
water and dust
tomer Services.
17