Beurer BM70: 9. Specifications

9. Specifications: Beurer BM70

the cuff tube is not properly inserted ( appears in the dis-

Memory 2 x 60 memory spaces

play),

Dimensions L 156 mm x W 117 mm x H 80 mm

inflation takes longer than 15 seconds ( appears in the dis-

play),

Weight Approx. 535 g (without batteries)

the inflation pressure is higher than 300 mmHg ( appears

Cuff size 24 to 36 cm

in the display),

Permissible operating

+10 °C to +40 °C, 40 85 % relative air

an error occurs when storing the measured values ( ap-

conditions

humidity (non-condensing)

pears in the display),

exceeding measurement range ( appears in the display).

Permissible storage

-10 °C to +60 °C, 10 – 90 % relative

conditions

air humidity, 800 1050 hPa ambient

In the about cases, you must repeat the measurement. Make

pressure

sure that the cuff tube is properly inserted and that you do not

Power supply

4 x 1,5 V

AA batteries

move or talk. Put the batteries back in if necessary, or else re-

place them.

Battery life For approx. 300 measurements,

depending on the blood pressure level

9. Specifications

and/or pump pressure

Model no. BM 70

Accessories Instruction for use, 4 x 1.5 V AA batter-

Measurement

Oscillometric, non-invasive blood

ies, storage pouch

method

pressure measurement on the upper

Classification Internal supply, IPX0, no AP or APG,

arm

continuous operation, type BF applied

Measurement range Cuff pressure 0 300 mmHg,

part

systolic 30 – 260 mmHg,

These specification are subject to change without notice for

diastolic 30 – 260 mmHg,

purpose of improvement.

Pulse 40 –199 beats/minute

Display accuracy Systolic ± 3 mmHg,

diastolic ± 3 mmHg,

pulse ± 5 % of the value shown

Measurement inac-

Max. permissible standard deviation

curacy

according to clinical testing:

systolic 8 mmHg /diastolic 8 mmHg

23