Beurer BM70: 9. Specifications

9. Specifications: Beurer BM70

•

the cuff tube is not properly inserted ( appears in the dis-

Memory 2 x 60 memory spaces

play),

Dimensions L 156 mm x W 117 mm x H 80 mm

•

inflation takes longer than 15 seconds ( appears in the dis-

play),

Weight Approx. 535 g (without batteries)

•

the inflation pressure is higher than 300 mmHg ( appears

Cuff size 24 to 36 cm

in the display),

Permissible operating

+10 °C to +40 °C, 40 – 85 % relative air

•

an error occurs when storing the measured values ( ap-

conditions

humidity (non-condensing)

pears in the display),

•

exceeding measurement range ( appears in the display).

Permissible storage

-10 °C to +60 °C, 10 – 90 % relative

conditions

air humidity, 800 – 1050 hPa ambient

In the about cases, you must repeat the measurement. Make

pressure

sure that the cuff tube is properly inserted and that you do not

Power supply

4 x 1,5 V

AA batteries

move or talk. Put the batteries back in if necessary, or else re-

place them.

Battery life For approx. 300 measurements,

depending on the blood pressure level

9. Specifications

and/or pump pressure

Model no. BM 70

Accessories Instruction for use, 4 x 1.5 V AA batter-

Measurement

Oscillometric, non-invasive blood

ies, storage pouch

method

pressure measurement on the upper

Classification Internal supply, IPX0, no AP or APG,

arm

continuous operation, type BF applied

Measurement range Cuff pressure 0 – 300 mmHg,

part

systolic 30 – 260 mmHg,

These specification are subject to change without notice for

diastolic 30 – 260 mmHg,

purpose of improvement.

Pulse 40 –199 beats/minute

Display accuracy Systolic ± 3 mmHg,

diastolic ± 3 mmHg,

pulse ± 5 % of the value shown

Measurement inac-

Max. permissible standard deviation

curacy

according to clinical testing:

systolic 8 mmHg /diastolic 8 mmHg

23