Beurer BM 60: 8. Rectifying faults 9. Specifications

8. Rectifying faults 9. Specifications: Beurer BM 60

•

Do not use detergents or solvents.

Display accuracy Systolic ± 3 mmHg, diastolic ± 3 mmHg,

•

On no account must you immerse the computer in water,

pulse ± 5 % of the value shown

otherwise liquid can enter it and cause demage.

Measurement

Max. permissible standard deviation

•

When storing the device, make sure that no heavy objects

inaccuracy

according to clinical testing:

are placed on top of it. Remove the batteries. The cuff tube

systolic 8 mmHg /diastolic 8 mmHg

should not have any sharp kinks.

Memory 2 x 60 memory spaces

8. Rectifying faults

Dimensions L 101 mm x W 130 mm x H 54 mm

E

rror messages can occur when

Weight Approx. 387 g (without batteries)

•

you move or talk during the measurement ( , ),

•

the cuff tube is not properly inserted ( ),

Cuff size 22 to 32 cm

•

when changing the batteries, the start/stop button was

Permissible operating

+10 °C to + 40 °C, ≤ 85 % relative

pressed,

conditions

air humidity (non-condensing),

•

inflation takes longer than 20 seconds ( ),

700 –1060 hPa ambient pressure

•

the inflation pressure is higher than 300 mmHg ( ).

Permissible storage

- 5 °C to + 50 °C, ≤ 85 % relative air

In the above cases, you must repeat the measurement. Make

and transport con-

humidity

sure that the cuff tube is properly inserted and that you do

ditions

not move or talk. Re-insert the batteries if necessary, or else

Power supply

4 x 1,5 V

AA batteries

replace them.

Battery life For approx. 300 measurements,

depending on the blood pressure level

9. Specifications

and/or pump pressure

Model no. BM 60

Accessories Instruction for use, 4 x 1.5 V AA batter-

Measurement

Oscillometric, non-invasive blood pres-

ies, storage pouch

method

sure measurement on the upper arm

Classification Internal supply, IPX0, no AP or APG,

Measurement range Cuff pressure 0 – 300 mmHg,

continuous operation, application part

systolic 50 – 250 mmHg,

type BF (cuff), CLASS II (when using a

diastolic 40 –180 mmHg,

mains part)

Pulse 40 –160 beats/minute

Maximum surface

Device: 44,1 °C

temperature

Mains part: 47,2 °C

23