Beurer BC 32: 8. Cleaning and Care 9. Technical details

8. Cleaning and Care 9. Technical details: Beurer BC 32

The cuff is fastened too tightly or loosely: E3;

Measurement

Max. permissible standard deviation

Errors occur during the measurement: E4;

inaccuracy

according to clinical testing:

The pump pressure is higher than 300 mmHg: E5;

systolic 8 mmHg /diastolic 8 mmHg

The batteries are almost empty E6.

Memory 2 x 60 memory spaces

In such cases, repeat the measurement. Ensure that you do not

move or speak. If necessary, reinsert or replace the batteries.

Dimensions L 72 mm x W 69 mm x H 29.2 mm

Weight Approx. 101 g (without batteries)

Cuff size 135 to 195 mm

8. Cleaning and Care

Permissible operating

+5 °C to +40 °C, 15-93 % relative air

Clean your blood pressure monitor carefully using a slightly

conditions

humidity (non-condensing)

damp cloth only.

Permissible storage

-25 °C to +70 °C, 10-93 % relative air

Do not use any detergents or solvents.

conditions

humidity, 700 –1060 hPa ambient pres-

Never hold the instrument under water as otherwise liquid

sure

can penetrate and damage the instrument.

Never place any heavy objects on the instrument.

Power supply

2 x 1.5 V

AAA batteries

Battery life For approx. 250 measurements,

9. Technical details

depending on the blood pressure level

Model no.

BC 32

and/or pump pressure

Type

BC32/1

Accessories Instruction for use, 2 x 1.5 V AAA batter-

Measurement

Oscillometric, non-invasive blood

ies, storage box

method

pressure measurement on the wrist

Classification Internal supply, IPX0, no AP or APG,

Measurement range Cuff pressure 0 300 mmHg,

continuous operation, type BF applied

systolic 50 – 250 mmHg,

part

diastolic 30 – 200 mmHg,

Technical information is subject to change without notification

Pulse 40 –180 beats/minute

to allow for updates.

Display accuracy Systolic ± 3 mmHg,

diastolic ± 3 mmHg,

pulse ± 5 % of the value shown