Beurer BC 40: 8. Cleaning and maintenance 9. Technical specifications

8. Cleaning and maintenance 9. Technical specifications: Beurer BC 40

you move or speak during the measurement ( appears

Display accu-

Systolic ±

3 mmHg, diastolic ± 3 mmHg, pulse

on the display)

racy

±

5 % of the value shown

the cuff is placed on the arm too tightly or loosely or if the

Measurement

Max. permissible standard deviation accord-

cuff is damaged ( appears on the display)

inaccuracy

ing to clinical testing: systolic 8 mmHg/dias-

the pump pressure is higher than 290 mmHg (

tolic 8 mmHg

appears on the display)

Memor

y 60 memory spaces

the batteries are almost worn out (

and )

.

Dimensions

L 76 mm x W 70 mm x H 23 mm

In such cases, repeat the measurement. If necessary, rein-

Weight Approximately 105 g (without batteries)

sert or replace the batteries.

Cuff size 12.5 to 21.5 cm

8. Cleaning and maintenance

Permissible

+ 10

°C to + 40 °C, ≤ 85 % relative humidity

Clean your blood pressure monitor carefully using a

operating con-

(non-condensing)

slightly damp cloth only.

ditions

Do not use any cleaning agents or solvents.

Permissible

- 20

°C to + 55 °C, ≤ 85 % relative humidity,

Under no circumstances should you hold the unit under

st

orage condi

-

800-1060

hPa ambient pressure

water, as this can cause liquid to enter and damage the unit.

tions

Never place any heavy objects on the device.

Power supply

2 x 1.5V

AAA batteries

9. Technical specifications

Battery life For approx. 300 measurements, depending on

Model no. BC 40

levels of blood pressure and pump pressure

Measurement

Oscillometric, non-invasive blood pressure

Accessories Instructions for use, 2 x 1.5V AAA batteries,

method

measurement on the wrist

storage box

Measurement

Cuff pressure 0 – 290 mmHg,

Classification Internal supply, continuous operation, applica

-

range

Systolic 50 – 250 mmHg,

tion part type BF, IPX0, no AP or APG

Diastolic 40 –180 mmHg,

Pulse 40 – 180 beats/minute

25