Hach-Lange SENSION+ MM374: инструкция

Раздел: Техника

Тип:

Инструкция к Hach-Lange SENSION+ MM374

DOC022.98.90253

sensION

+ MM374

12/2013, Edition 4

User Manual

Bedienungsanleitung

Manuale dell'utente

Manuel d'utilisation

Manual del usuario

Manual do utilizador

Uživatelská příručka

Brugsanvisning

Gebruikershandleiding

Instrukcja obsługi

Bruksanvisning

Käyttäjän käsikirja

Ръководство на потребителя

Használati útmutató

Manual de utilizare

Naudotojo vadovas

Руководство пользователя

Kullanım Kılavuzu

Návod na obsluhu

Navodila za uporabo

Korisnički priručnik

Εγχειρίδιο χρήσης

Kasutusjuhend

English...................................................................................................................................................................................................3

Deutsch...............................................................................................................................................................................................23

Italiano.................................................................................................................................................................................................44

Français..............................................................................................................................................................................................65

Español...............................................................................................................................................................................................86

Português........................................................................................................................................................................................107

Čeština..............................................................................................................................................................................................128

Dansk.................................................................................................................................................................................................149

Nederlands......................................................................................................................................................................................169

Polski.................................................................................................................................................................................................190

Svenska............................................................................................................................................................................................211

Suomi.................................................................................................................................................................................................231

български.......................................................................................................................................................................................252

Magyar..............................................................................................................................................................................................274

Română............................................................................................................................................................................................295

lietuvių kalba...................................................................................................................................................................................316

Русский............................................................................................................................................................................................337

Türkçe................................................................................................................................................................................................360

Slovenský jazyk............................................................................................................................................................................381

Slovenski..........................................................................................................................................................................................402

Hrvatski.............................................................................................................................................................................................423

Ελληνικά...........................................................................................................................................................................................444

eesti keel..........................................................................................................................................................................................467

2

Table of contents

Specification Details

Specifications on page 3 Advanced operation on page 12

Resolution pH: 0.1/0.01/0.001, ORP: 0.1/1 mV, ISE:

programmable, temperature: 0.1 ºC (0.18 ºF), EC:

General information on page 4 Data logger on page 13

variable, resistivity: variable, NaCl: variable, TDS:

variable

Installation on page 5 Maintenance on page 15

Measuring error (±

pH: ≤ 0.005, ORP: ≤ 0.2 mV, temperature: ≤ 0.2 ºC (≤

User interface and navigation

Troubleshooting on page 17

1 digit)

0.36 ºF), EC: ≤ 0.5%, resistivity: ≤ 0.5 %, NaCl: ≤

on page 8

0.5 %, TDS: ≤ 0.5%

Startup on page 9 Replacement parts and accessories

Reproducibility (±

pH: ± 0.001, ORP: ± 0.1 mV, temperature: ± 0.1 ºC (±

on page 18

1 digit)

0.18 ºF), EC: ± 0.1%, resistivity: ± 0.1%, NaCl ± 0.1%,

TDS ± 0.1%

Standard operation on page 9 Standard solutions on page 19

Data storage 330 results and last 9 calibrations

Specifications

Connections 2 Combined or indicator probe: BNC connector (Imp.

>10

12

Ω); 2 Reference electrodes: banana connector;

Specifications are subject to change without notice.

A.T.C. type Pt 1000: banana or telephonic; 2 magnetic

stirrers: RCA connector

Specification Details

Conductivity probe with built-in Pt1000 sensor (or NTC

Dimensions 35 x 20 x 11 cm (13.78 x 7.87 x 4.33 in.)

10 kΩ probe): telephonic connector

RS232C for printer or PC: telephonic connector;

Weight 1100 g (2.43 lb)

external PC keyboard: mini DIN connector

Meter enclosure IP42

Temperature correction Channel 1 pH: Pt 1000 temperature probe (A.T.C.),

NTC 10 kΩ probe, manual, Channel 2 temperature,

Power requirements

100–240 V, 0.4 A, 47-63 Hz

isopotential pH programmable, standard value 7.00,

(external)

CE: Pt 1000 temperature probe (A.T.C.), linear

Meter protection class Class II

function, TC=0.00 to 9.99%/Temperature, TRef: 20°C

(68 °F) or 25°C (77 °F), non-linear function for natural

Pollution degree 2

waters (UNE EN 2788) Channel 2 pH: Pt

1000 temperature probe (A.T.C.), manual, NTC 10KΩ

Installation category Category II

probe, isopotential pH programmable, standard value

7.00

Altitude requirements Standard 2000 m (6562 ft) ASL (Above Sea Level)

Measurement display

Continuous measurement, by stability, by time

Storage temperature –15 to +65 °C (5 to +149 °F)

lock

Operating temperature 0 to 40 °C (41 to 104 °F)

Display Liquid crystal, backlit, 128 x 64 dots

Operating humidity < 80% (non-condensing)

Keyboard PET with protective treatment

Certification CE

English 3

General information

Precautionary labels

Revised editions are found on the manufacturer’s website.

Read all labels and tags attached to the instrument. Personal injury or

damage to the instrument could occur if not observed. A symbol on the

Safety information

instrument is referenced in the manual with a precautionary statement.

N O T I C E

This symbol, if noted on the instrument, references the instruction

manual for operation and/or safety information.

The manufacturer is not responsible for any damages due to misapplication or

misuse of this product including, without limitation, direct, incidental and

consequential damages, and disclaims such damages to the full extent permitted

Electrical equipment marked with this symbol may not be disposed of

under applicable law. The user is solely responsible to identify critical application

in European domestic or public disposal systems. Return old or end-

risks and install appropriate mechanisms to protect processes during a possible

of-life equipment to the manufacturer for disposal at no charge to the

equipment malfunction.

user.

Please read this entire manual before unpacking, setting up or operating

this equipment. Pay attention to all danger and caution statements.

Failure to do so could result in serious injury to the operator or damage

to the equipment.

Product overview

Make sure that the protection provided by this equipment is not impaired.

The sensION

+ meters are used with probes to measure various

Do not use or install this equipment in any manner other than that

parameters in water.

specified in this manual.

The sensION

+ MM374 meter has two measurement channels to

measure pH, ORP (mV), conductivity or ISE (concentration) with a

Use of hazard information

selective probe. Channel 1 measures one or two parameter individual or

simultaneously and up to two probes can be connected. Channel

D A N G E R

2 measures pH, ORP (mV) or ISE. Measurement data can be stored and

Indicates a potentially or imminently hazardous situation which, if not avoided, will

transferred to a printer or PC.

result in death or serious injury.

Product components

W A R N I N G

Refer to Figure 1 to make sure that all components have been received.

Indicates a potentially or imminently hazardous situation which, if not avoided,

If any items are missing or damaged, contact the manufacturer or a

could result in death or serious injury.

sales representative immediately.

C A U T I O N

Indicates a potentially hazardous situation that may result in minor or moderate

injury.

N O T I C E

Indicates a situation which, if not avoided, may cause damage to the instrument.

Information that requires special emphasis.

4

English

Figure 1 Meter components

Installation

Assemble the probe holder

Follow the numbered steps to assemble the probe holder and to connect

the magnetic stirrer.

1 Electrolyte for the probe 6 Power supply

2 Buffer solutions (pH 4.01 and pH

7 Probe holder

7.00)

3 Calibration beakers (with magnetic

8 Rod

bar inside)

4 Standard solution (1413 µS/cm) 9 Probes (included with kits only)

5 Calibration beaker (with magnetic

10 Meter

bar inside)

English 5

1 2

3 4

6 English

Connect to AC power

Figure 2 Connector panel

D A N G E R

Electrocution hazard. If this equipment is used outdoors or in

potentially wet locations, a Ground Fault Circuit Interrupt (GFCI/GFI)

device must be used to connect the equipment to its main power

source.

The meter can be powered by AC power with the universal power

adapter.

1. Select the correct adapter plug for the power outlet from the adapter

kit.

2. Connect the universal power adapter to the meter (Figure 2).

3. Connect the universal power adapter to an AC receptacle (Figure 3).

4. Turn the meter on.

1 Magnetic stirrer 1 connector,

7 PC keyboard, mini DIN connector

Channel 1

2 Reference electrode (separated

8 Conductivity probe connector,

electrodes) connector, Channel 1

Channel 2

3 Reference electrode (separated

9 Temperature probe connector,

electrodes) connector, Channel 2

Channel 2

4 Separated temperature probe

10 Combined pH electrode (or

connector, Channel 2

indicator) connector, Channel 2

5 RS-232 for printer or PC connector 11 Combined pH electrode (or

indicator) connector, Channel 1

6 Power supply 12 Magnetic stirrer 2 connector,

Channel 1

English 7

Figure 3 AC power connection

User interface and navigation

User interface

Keypad description

1 RETURN key: cancel or exit the

5 DOWN key: scroll to other options,

current menu screen to the

change a value

previous menu screen

Note: Position the equipment so it is not difficult to disconnect power to the

2 MEASUREMENT key: confirm the

6 ON/OFF: turn on or turn off the

equipment.

selected option

meter

3 UP key: scroll to other options,

7 LEFT key: change between

change a value

Channel 1 and Channel 2, enter

numbers and letters

4 RIGHT key: change between

Channel 1 and Channel 2, enter

numbers and letters

Display description

The meter display shows the concentration, units, temperature,

calibration status, operator ID, sample ID, date and time.

8

English

Figure 4 Single screen display

Change the language

The display language is selected when the meter is powered on for the

first time.

1. Use the or to select a language from a list.

2. Confirm with . The measurement screen shows DATA OUTPUT.

3. Select Deactivated if no printer or PC is connected and confirm.

Refer to Select the data output on page 12 for more information

1 Sample ID 5 Measurement channel

about Data Output.

2 Measurement unit and value

6 Sample temperature (ºC or ºF)

(conductivity or ISE)

Standard operation

3 Measurement unit and value (pH,

7 Visual measurement timer

ORP/mV)

Calibration

4 Measurement mode or time and

D A N G E R

date

Chemical exposure hazard. Obey laboratory safety procedures and

Navigation

wear all of the personal protective equipment appropriate to the

Use the to return to the previous menu. Use the measure key to

chemicals that are handled. Refer to the current safety data sheets

(MSDS/SDS) for safety protocols.

take a sample measurement or to confirm options. Use the arrow keys

to scroll to other options or to change a value. To change the

parameters use the arrow keys and . Refer to each task for specific

Calibration settings

instructions.

The calibration settings contain Calibration type, Calibration frequency

and Display options.

Startup

1. From the main menu use the or to select CALIBRATION.

Turn the meter on and off

Confirm.

2. Use the to enter the calibration menu.

N O T I C E

3. Use the or to select the following options:

Make sure that the probe is connected to the meter before the meter is turned on.

Option Description

Push to turn on or turn off the meter. If the meter does not turn on,

make sure that the AC power supply is properly connected to an

Stability C.: Criterion by stability—select Fast, Standard or Strict.

electrical outlet.

English 9

Option Description

Option Description

Calibration

pH: Calibration type—select Technical buffers,

Theoretical calibration pH: The probe calibration data is replaced at

type

DIN19266 Buffers, User Buffers, To a X value, data

25 °C (77 °F). EC: The probe calibration data is

introduction or Theoretical calibration.EC: Calibration type

replaced with C=1.000 cm

-1

—select Molar Standards, Demal Standards, NaCl

Standards, Calibration to a X value, data introduction or

Molar Standards 147 µS/cm, 1413 µS/cm, 12.88 mS/cm and

Theoretical calibration. Refer to Calibration types for more

111.8 mS/cm at 25 °C (77 °F)

information.

Demal Standards 1049 µS/cm, 12.85 mS/cm and 111.31 mS/cm at

Cal. frequency Calibration reminder—can be set between 0–7 days

25 °C (77 °F)

(default daily). The display shows the remaining time to

NaCl Standards 1014.9 µS/cm 25 °C (77 °F)

the new calibration. Refer to Set the calibration reminder

on page 11 for more information.

Calibration procedure

Display mV Display mV—select YES or NO to show mV.

This procedure is for general use with liquid calibration solutions. Refer

to the documents that are included with each probe for additional

Calibration types

information.

Different calibration types can be selected.

Note: Solutions should be stirred during calibration. For more information about the

stirring settings, refer to Change the stirring settings on page 13.

1. From the main menu use the or to select CALIBRATION.

Note: If Channel 1 is used as a double channel (pH and EC) the temperature will

Confirm.

be measured by the conductivity cell with integrated ATC. The conductivity cell

2. Use the to enter the calibration menu.

must be immersed in the pH buffer solution as well during the pH calibration.The

manual temperature modification is not possible.

3. Use the or to select Calibration type.

Option Description

1. Pour the buffers or calibration solutions into the labeled calibration

beakers.

Technical buffers pH 2.00, 4.01, 7.00, 9.21 and 10.01 at 25 °C

(77 °F)

2. From the main menu use the or and and to select the

CALIBRATION parameter. Confirm.

DIN19266 Buffers pH 1.679, 4.006, 6.865, 7.000, 9.180, 10.012 and

12.454

3. If required select the Operator ID (1 to 10) and confirm.

4. Rinse the probe with deionized water and put the probe into the first

User Buffers Selected when the technical or DIN19266 buffers

calibration beaker. Be sure that there are no air bubbles in the

are not used. Refer to Technical buffer solutions

(DIN 19267) on page 19 for pH values of

membrane.

specific buffer sets at varying temperatures.

5. Push to Start calibration.

Calibration to a X value To adjust manually any scale value of the

6. Push to measure the first calibration solution.

measured pH or conductivity.

The next calibration solution is shown.

Data introduction Manual probe constant introduction.

7. Rinse the probe with deionized water and put the probe into the

second calibration beaker. Be sure that there are no air bubbles in

the membrane.

10

English

8. Push to measure the second calibration solution.

4. Use the and to advance to the next step and use the or to

The next calibration solution is shown.

change a value. Confirm.

Push to start the calibration.

9. Rinse the probe with deionized water and put the probe into the third

calibration beaker. Be sure that there are no air bubbles in the

membrane.

Sample measurements

10. Push to measure the third calibration solution.

Each probe has specific preparation steps and procedures for taking

When the calibration is good, the display briefly shows Calibration

sample measurements.

OK and then returns to the main menu.

Note: When a printer is connected the print menu opens and the result can be

1. From the main menu use the or and and to select

printed.

MEASURE. Confirm.

2. Use the to change the following settings. Confirm every entry.

View the calibration data

Data from the most recent calibration can be shown.

Option Description

Resolution Select the resolution: 1, 0.1, 0.01 (default) or 0.001

1. From the main menu use the or to select DATA LOGGER.

Confirm.

Measure Stability—select By stability Criterion: Fast (variation <

2. Select Display data.

0.02 pH in 6 s), Standard (variation < 0.01 pH in 6 s) or Strict

(variation < 0.002 pH in 6 s). In continuous—enter the time

3. Select Calibration data and confirm with . The last calibration data

interval for the In continuous Acquis. interval (data storage or

is shown.

printing data). By time—enter the time interval for data storage

or printing data.

pH-the slope and offset values are shown alternating with the

deviation (in %) and calibration temperature.

Display mV Display mV—select YES or NO to show mV.

ORP-the measured mV value and calibration temperature are

Limits Limits—select YES or NO. YES: Enter the upper and lower

shown.

limits. An acoustic warning appears when the measurement is

Conductivity-the cell constant and calibration temperature for each

out of limit. The report output shows an A next to the

measured value when the measurement was out of limit.

standard are shown.

Isopotential Isopotential—change the Isopotential pH value in Data

Set the calibration reminder

introduction. Select Calculate to calibrate the probe again.

pH: The calibration reminder can be set between 0 to 23 hours or

TC TC—select Linear or Natural waters. Linear: enter a value in

1-7 days (default 1 day). EC: The calibration reminder can be set

%/°C (default 2.00%/Temperature). Natural waters: Non-linear

between 0 to 99 days (default 15 days). The display shows the

for natural waters, according to EN27888

remaining time to the new calibration.

Note: When 0 days is selected, the calibration reminder is turned off.

Tref Reference temperature—select between 20 or 25 °C or Other

temperature.

1. From the main menu use the or to select CALIBRATION.

3. Push to start the measurement.

Confirm.

Note: If the measurement is not stabilizing after 120 seconds, the meter turns

2. Use the to enter the calibration menu.

automatically into the continuous measurement mode.

3. Use the or to select Cal. frequency and confirm.

English

11

Advanced operation

1. From the main menu use the or to select SYSTEM. Confirm.

Change the measurement units

2. Use the or to select Data Output and confirm.

3. Use the or to select

The measurement units can be changed individually for each channel.

Option Description

1. From the main menu use the or to select SYSTEM. Confirm.

Deactivated Select Deactivated if no printer or PC is connected.

2. Use the or to select Measurement units and confirm.

3. Select Channel 1 and confirm.

For Printer Select Dot matrix printer or Thermal printer.

4. Select Parameter 1 and then Parameter 2 and select one of the

For Computer Select Terminal, LabCom or LabCom Easy. The LabCom

following options:

Software controls several modules, pH and conductivity

meters, automatic burettes, Samplers and so on from a

Option Description

computer. The LabCom Easy software gets measurements

and calibration data from the meters

Parameter 1 mV, pH, ISE or Disabled

Parameter 2 EC, NaCl, TDS, Ω or Disabled

Change the date and time

5. Select Channel 2 and confirm with . Select mV, pH or ISE and

The date and time can be changed from the Date / Time menu.

confirm.

1. From the main menu use the or to select SYSTEM. Confirm.

Use a sample ID

2. Use the or to select Date / Time and confirm.

The sample ID tag is used to associate readings with a particular sample

3. Use the and to advance to the next step and use the or to

location. If assigned, stored data will include this ID.

change a value. Confirm.

The current date and time will be shown on the display.

1. From the main menu use the or to select SYSTEM. Confirm.

2. Use the or to select Sample ID and confirm.

Adjust the display contrast

3. Use the or to select

1. From the main menu use the or to select SYSTEM. Confirm.

Option Description

2. Use the or to select Display contrast and confirm.

Automatic A consecutive number will be automatically assigned to every

3. Use the and to adjust the contrast of the display and confirm.

sample.

Manual A keyboard or a barcode scanner is required to enter the

Adjust the temperature

sample ID name before taking a measurement (maximum

15 characters).

The temperature measurement can be adjusted at 25 °C (77 °F) and/or

85 °C (185 °F) to increase accuracy.

Select the data output

Data can be stored or transferred to a printer or to a PC.

12

English

4. When the stirrer is turned on, use the or to change the stirring

1. Put the probe and a reference thermometer in a container of water at

speed in %.

approximately 25 °C and allow the temperature to stabilize.

Note: Use the or to change the stirring speed during calibration and

2. Compare the temperature read by the meter with that of the

during a measurement.

reference thermometer. The difference is the adjustment value for

the meter.

Turn the stirrer on or off

Example: reference thermometer: 24.5 °C; meter: 24.3 °C.

The stirrer 1 works with Channel 1 and 2 (Stirrer 1). A second stirrer can

Adjustment value: 0.2 °C.

be connected to channel 2 (Stirrer 2). For the stirrer 2 activation refer to

3. Enter the adjustment value for the 25 °C reading:

the following steps.

a. From the main menu use the or to select SYSTEM. Confirm.

1. From the main menu use the or to select SYSTEM. Confirm

b. Use the or to select Readjust temp. and confirm.

with .

c. Select Channel 1 or Channel 2 and confirm. The temperature for

2. Use the or to select Stirrer N.2 and confirm with .

channel 1 is measured by the conductivity cell and channel 2 is

3. Use the or to select YES to turn the stirrer 2 on.

measured by the pH probe. If no Conductivity cell is connected,

Note: Select NO to turn the stirrer 2 off.

the temperature value must be entered manually or the

temperature measured in channel 2 can be applied to the meter.

Change the temperature units

d. Use the or to select 25 °C and confirm.

e. Use the arrow keys to enter the adjustment value for 25 °C.

The temperature units can be changed to Celsius or Fahrenheit.

Confirm.

1. From the main menu use the or to select SYSTEM. Confirm.

4. Put the probe and a reference thermometer in a container of water at

approximately 85 °C and allow the temperature to stabilize.

2. Use the or to select Temperature units and confirm.

5. Compare the temperature from the meter with that of the reference

3. Use the or to select between Celsius or Fahrenheit and confirm.

thermometer. The difference is the adjustment value for the meter.

Data logger

a. Use the or to select 85 °C and confirm.

b. Use the arrow keys to enter the adjustment value for 85 °C.

Display data

Confirm.

c. Select Save changes and confirm.

The Display data log contains Measurement data, Electrode report and

Calibration data. The stored data can be sent to a printer or to a PC.

Change the stirring settings

When the data log becomes full (400 data points), the oldest data point

is deleted when a new data point is added.

The magnetic stirrer can be turned on and the stirring speed can be

changed in the Stirring menu.

1. From the main menu use the or to select SYSTEM. Confirm.

2. Use the or to select Stirring and confirm.

3. To turn on/off the stirrer push .

English

13

1. From the main menu use the or to select DATA LOGGER and

data, Electrode data, Calibration data, Calibration report or

confirm.

Instrument condit.

2. Use the or to select Display data and confirm.

Report output

3. Use the or to select

Option Description

N O T I C E

Measurement data Measurement data—stores automatically each time a

The data output (printer or PC) needs to be selected first, so that the Type of

sample is measured

report menu is available (refer to Select the data output on page 12).

Electrode report Electrode report—stores automatically the electrode

Different report output types can be selected when a printer or a PC is

history and measurement conditions

connected.

Calibration data Calibration data—stores automatically the current

1. From the main menu use the or to select SYSTEM. Confirm.

calibration

2. Use the or to select Type of report and confirm.

Delete data

3. When a printer or a computer and Terminal is connected, use the

or to select

The entire measurement data or electrode report log can be deleted to

remove data that has already been sent to a printer or PC.

Option Description

Reduced Select Several or One sample as an output format

1. From the main menu use the or to select DATA LOGGER and

confirm.

2. Use the or to select Erase and confirm.

3. Use the or to select Measurement data or Electrode report and

confirm. Confirm again to delete the data.

The entire log is deleted at once.

Send data to a printer or to a computer

N O T I C E

The data output (printer or PC) needs to be selected first, so that the Print menu

is available (refer to Select the data output on page 12).

Note: Refer to Report output on page 14 to select the report output type.

1. From the main menu use the or to select DATA LOGGER and

confirm.

2. Use the or to select Print and confirm. Select one of the

following options and confirm with to print the data: Measurement

14 English

Option Description

C A U T I O N

Standard Select Several or One sample as an output format. Select

Personal injury hazard. Only qualified personnel should conduct the tasks

several: Users: The user name appears on the printed report

described in this section of the manual.

(17 characters). Header:The company name can be added as a

header (40 characters) and appears on the print report. Identify

Clean the instrument

sensor:The sensor model and the sensor serial number can be

added and appears on the print report.

N O T I C E

GLP Select Several or One sample as an output format. Select

Never use cleaning agents such as turpentine, acetone or similar products to

several: Users: The user name appears on the printed report

clean the instrument including the display and accessories.

(17 characters). Header:The company name can be added as a

header (40 characters) and appears on the print report. Identify

Clean the exterior of the instrument with a moist cloth and a mild soap

sensor:The sensor model and the sensor serial number can be

solution.

added and appears on the print report.

4. When a computer is connected and LabCom Easy (refer to Select

Clean the probe

the data output on page 12 for more information) is selected, use the

Clean the probe as needed. Refer to Troubleshooting on page 17 for

or to select

more information about cleaning. Refer to the probe documentation for

Option Description

information about the probe maintenance.

Users The user name appears on the printed report

Replace the magnetic stirrer

(17 characters).

If the magnetic stirrer does not start, follow the numbered steps to

Identify sensor The sensor model and the sensor serial number can be

added and appears on the print report.

replace the magnetic stirrer.

Maintenance

W A R N I N G

Multiple hazards. Do not disassemble the instrument for maintenance or service.

If the internal components must be cleaned or repaired, contact the

manufacturer.

English 15

Аннотация для Hach-Lange SENSION+ MM374 в формате PDF