Hach-Lange POLYMETRON 9500 Ultrapure pH_ORP Module User Manual: инструкция

Раздел: Техника

Тип:

Инструкция к Hach-Lange POLYMETRON 9500 Ultrapure pH_ORP Module User Manual

DOC023.98.93063

Ultrapure pH/ORP Module

10/2013, Edition 3

User Manual

Bedienungsanleitung

Manuale d'uso

Manuel d'utilisation

Manual de usuario

Manual do utilizador

用户手册

Gebruikershandleiding

Instrukcja obsługi

Käyttäjän käsikirja

Руководство пользователя

English..............................................................................................................................3

Deutsch..........................................................................................................................18

Italiano............................................................................................................................34

Français.........................................................................................................................49

Español..........................................................................................................................64

Português......................................................................................................................80

中文.................................................................................................................................95

Nederlands.................................................................................................................109

Polski............................................................................................................................125

Suomi............................................................................................................................141

Русский........................................................................................................................156

2

Specifications

Specifications are subject to change without notice.

pH/ORP sensors

8350.3 8350.4 8350.5 8351

Measuring pH in

Measuring pH in a

Measuring pH at

Application

samples containing

waste water

Measuring ORP

high temperatures

hydrofluoric acid

environment

Material PPS PPS CPVC PPS

Measurement range 0—12 pH 0—14 pH 0—12 pH ± 1500 mV

Maximum

110 °C 110 °C 80 °C 110 °C

temperature

Maximum pressure 10 bar 10 bar 10 bar 10 bar

Repeatability (week) < 0.05 pH < 0.05 pH < 0.05 pH < 0.05 pH

Offset @ pH7 ± 0.34 pH ± 0.34 pH ± 0.34 pH N/A

Slope 56—61 mV/pH 56—61 mV/pH 56—61 mV/pH N/A

Reference

< 50 Kohm < 50 Kohm < 50 Kohm < 50 Kohm

impedance @ 25 °C

Glass impedance @

100—150 Mohm 150—500 Mohm 50—250 Mohm N/A

25 °C

8362 8417

Measuring pH in pure or ultra pure

Measuring pH in industrial and

Application

water

waste water treatment plants

Material 316L stainless steel Glass membrane, ceramic junction

Measurement range 2—12 pH 0—14 pH

Maximum temperature 80 °C 110 °C

16 bar @ 25 °C;

Maximum pressure

10 bar @ 25 °C

6 bar @ 100 °C

Repeatability (24 hours) < 0.01 pH 0.02 pH

6

Input impedance > 10

Mohm 100 Mohm

Antimony electrode sensors

8346 / 8347

Application Measuring pH or ORP in dirty water

Material Polypropylene

80 °C (pH measurements with 8346)

Maximum temperature

120 °C (pH measurements with 8347)

50 °C (ORP measurements)

Sample pressure Atmospheric

Maximum pressure (electrolyte container) 5 bar

English 3

Аннотация для Hach-Lange POLYMETRON 9500 Ultrapure pH_ORP Module User Manual в формате PDF