Hach-Lange POLYMETRON 9500 Ultrapure pH_ORP Module User Manual: инструкция
Раздел: Техника
Тип:
Инструкция к Hach-Lange POLYMETRON 9500 Ultrapure pH_ORP Module User Manual

DOC023.98.93063
Ultrapure pH/ORP Module
10/2013, Edition 3
User Manual
Bedienungsanleitung
Manuale d'uso
Manuel d'utilisation
Manual de usuario
Manual do utilizador
用户手册
Gebruikershandleiding
Instrukcja obsługi
Käyttäjän käsikirja
Руководство пользователя

English..............................................................................................................................3
Deutsch..........................................................................................................................18
Italiano............................................................................................................................34
Français.........................................................................................................................49
Español..........................................................................................................................64
Português......................................................................................................................80
中文.................................................................................................................................95
Nederlands.................................................................................................................109
Polski............................................................................................................................125
Suomi............................................................................................................................141
Русский........................................................................................................................156
2

Specifications
Specifications are subject to change without notice.
pH/ORP sensors
8350.3 8350.4 8350.5 8351
Measuring pH in
Measuring pH in a
Measuring pH at
Application
samples containing
waste water
Measuring ORP
high temperatures
hydrofluoric acid
environment
Material PPS PPS CPVC PPS
Measurement range 0—12 pH 0—14 pH 0—12 pH ± 1500 mV
Maximum
110 °C 110 °C 80 °C 110 °C
temperature
Maximum pressure 10 bar 10 bar 10 bar 10 bar
Repeatability (week) < 0.05 pH < 0.05 pH < 0.05 pH < 0.05 pH
Offset @ pH7 ± 0.34 pH ± 0.34 pH ± 0.34 pH N/A
Slope 56—61 mV/pH 56—61 mV/pH 56—61 mV/pH N/A
Reference
< 50 Kohm < 50 Kohm < 50 Kohm < 50 Kohm
impedance @ 25 °C
Glass impedance @
100—150 Mohm 150—500 Mohm 50—250 Mohm N/A
25 °C
8362 8417
Measuring pH in pure or ultra pure
Measuring pH in industrial and
Application
water
waste water treatment plants
Material 316L stainless steel Glass membrane, ceramic junction
Measurement range 2—12 pH 0—14 pH
Maximum temperature 80 °C 110 °C
16 bar @ 25 °C;
Maximum pressure
10 bar @ 25 °C
6 bar @ 100 °C
Repeatability (24 hours) < 0.01 pH 0.02 pH
6
Input impedance > 10
Mohm 100 Mohm
Antimony electrode sensors
8346 / 8347
Application Measuring pH or ORP in dirty water
Material Polypropylene
80 °C (pH measurements with 8346)
Maximum temperature
120 °C (pH measurements with 8347)
50 °C (ORP measurements)
Sample pressure Atmospheric
Maximum pressure (electrolyte container) 5 bar