Hach-Lange POLYMETRON 9500 Conductivity Module User Manual: инструкция

Раздел: Техника

Тип:

Инструкция к Hach-Lange POLYMETRON 9500 Conductivity Module User Manual

DOC023.98.93060

Ultrapure Conductivity Module

10/2013, Edition 3

User Manual

Benutzerhandbuch

Manuale d'uso

Manuel d'utilisation

Manual de usuario

Manual do utilizador

用户手册

Gebruikershandleiding

Instrukcja obsługi

Käyttäjän käsikirja

Руководство пользователя

English..............................................................................................................................3

Deutsch..........................................................................................................................22

Italiano............................................................................................................................43

Français.........................................................................................................................63

Español..........................................................................................................................83

Português....................................................................................................................105

中文...............................................................................................................................126

Nederlands.................................................................................................................143

Polski............................................................................................................................164

Suomi............................................................................................................................184

Русский........................................................................................................................204

2

Specifications

Specifications are subject to change without notice.

Contacting conductivity sensors

Specification 8310 8311 8312

Drinking water and wastewater treatment

Application

Chemical processes

Demineralized and softened water

Body material Black PSU Black PSU Black PSU

Internal electrode SS 316L SS 316L Graphite

External electrode SS 316L SS 316L Graphite

Insulator PSU PSU PSU

Connector Glass polyester (IP65) Glass polyester (IP65) Glass polyester (IP65)

-1

-1

-1

Cell constant K 0.01 (cm

) 0.1 (cm

) 1.0 (cm

)

Accuracy < 2% < 2% < 2%

-1

-1

-1

Measurement range 0.01—200 μS.cm

0.1 μS—2 mS.cm

1 μS—20 mS.cm

Temperature response < 30 seconds < 30 seconds < 30 seconds

Maximum temperature 125 °C 125 °C 125 °C

Maximum pressure 10 bars 10 bars 10 bars

Sample connection ¾ inch NPT ¾ inch NPT ¾ inch NPT

Specification 8315 8316 8317 8394

Monitoring of

ultrapure water in

Pure water production

pharmaceutical and

Application

Process water monitoring (condensates, cleaning cycles, heat

food industries

exchangers)

Suitable for CIP-SIP

processes

Body material SS 316L SS 316L SS 316L SS 316L

Internal electrode SS 316L SS 316L Graphite SS 316L

External electrode SS 316L SS 316L Graphite SS 316L

Insulator PES PES PES PEEK

Connector Glass polyester

Glass polyester

Glass polyester

Glass polyester

(IP65)

(IP65)

(IP65)

(IP65)

-1

-1

-1

-1

Cell constant K 0.01 (cm

) 0.1 (cm

) 1.0 (cm

) 0.01 (cm

)

Accuracy < 2% < 2% < 2% < 2%

-1

-1

-1

-1

Measurement range 0.01—200 μS.cm

0.1 μS—2 mS.cm

1 μS—20 mS.cm

0.01—200 μS.cm

Temperature

< 30 seconds < 30 seconds < 30 seconds < 30 seconds

response

Maximum

150 °C 150 °C 150 °C 150 °C

temperature

English 3