Beurer BM 26: ENGLISH
ENGLISH: Beurer BM 26
ENGLISH
Dear Customer,
2. Important information
Thank you for choosing one of our products. Our name stands
Signs and symbols
for high-quality, thoroughly tested products for the applications
The following symbols are used in these instructions for use, on
in the areas of heat, weight, blood pressure, body temperature,
the packaging and on the type plate for the device and acces-
pulse, gentle therapy, massage and air.
sories:
Please read these instructions for use carefully and keep them
for later use, be sure to make them accessible to other users
Caution
and observe the information they contain.
Best regards,
Your Beurer Team
Note
Note on important information
1. Getting to know your instrument
The upper arm blood pressure monitor is used for non-invasive
Follow instructions for use
measurement and monitoring of adults’ arterial blood pressure.
You can use it to measure your blood pressure quickly and
easily, storing the results and displaying the progression of
Type BF applied part
readings together with the average.
A warning is issued for anyone suffering from cardiac arrhyth-
Direct current
mia.
Disposal in accordance with EC Directive
The values determined are classified and graphically evaluated
2002/96/EC – WEEE (Waste Electrical and
according to WHO guidelines.
Electronic Equipment)
Keep these instructions carefully for further use and also let
other users have access to them.
Manufacturer
Storage
Permissible storage temperature and
55°C
humidity
-20°C
RH ≤85%
12
curacy. The same also applies to very low blood pressure,
Operating
Permissible operating temperature and
40°C
diabetes, circulatory disorders and arrhythmias as well as
humidity
10°C
RH ≤85%
chills or shaking.
•
The blood pressure monitor must not be used in connection
Keep dry
with a high-frequency surgical unit.
•
Only use the device on people who have the specified upper
arm measurement for the device.
SN Serial number
•
Please note that when inflating, the functions of the limb in
The CE labelling certifies that the product
question may be impaired.
complies with the essential requirements of
0483
•
During the blood pressure measurement, blood circulation
Directive 93/42/EEC on medical products
must not be stopped for an unnecessarily long time. If the
device malfunctions, remove the cuff from the arm.
Advice on use
•
Avoid any mechanical restriction, compression or bending of
•
In order to ensure comparable values, always measure your
the cuff line.
blood pressure at the same time of day.
•
Do not allow sustained pressure in the cuff or frequent meas-
•
Before every measurement, relax for about five minutes.
urements. The resulting restriction of the blood flow may
•
If you want to perform several measurements on the same
cause injury.
person, wait five minutes between each measurement.
•
Ensure that the cuff is not placed on an arm in which the
•
Do not take a measurement within 30 minutes after eating,
arteries or veins are undergoing medical treatment, e.g. intra-
drinking, smoking or exercising.
vascular access or therapy, or an arteriovenous (AV) shunt.
•
Repeat the measurement if you are unsure of the measured
•
Do not use the cuff on people who have undergone a mas-
value.
tectomy.
•
The measurements taken by you are for your information
•
Do not place the cuff over wounds as this may cause further
only – they are not a substitute for a medical examination!
injury.
Discuss the measurements with your doctor, and never base
•
The blood pressure monitor can only be operated with bat-
any medical decisions on them (e.g. medicines and their
teries.
administration)!
•
To conserve the batteries, the monitor switches off automati-
•
Do not use the blood pressure monitor on newborns, preg-
cally if no buttons are pressed for 3 minutes.
nant women or patients with preeclampsia.
•
The device is only intended for the purpose described in
•
Cardiovascular diseases may lead to incorrect measure-
these instructions for use. The manufacturer is not liable for
ments or have a detrimental effect on measurement ac-
damage resulting from improper or careless use.
13
•
Never use different types of battery, battery brands or bat-
Storage and Care
teries with different capacities. You should preferably use
•
The blood pressure monitor is made up of precision electron-
alkaline batteries.
ic components. Accuracy of readings and the instrument’s
service life depend on careful handling.
Repair and disposal
– You should protect the device from impact, moisture, dirt,
•
Batteries do not belong in domestic refuse. Used batteries
major temperature fluctuations and direct exposure to the
should be disposed of at the collection points provided.
sun’s rays.
•
Never open the instrument. If these instructions are not
– Never drop the device.
heeded, the warranty will be null and void.
– Do not use near strong electromagnetic fields, i.e. keep it
•
Never attempt to repair the instrument or adjust it yourself.
away from any radio systems and mobile phones.
We can no longer guarantee perfect functioning if you do.
– Only ever use the cuffs provided with the monitor or origi-
•
Repairs may only be performed by Customer Service or
nal replacement cuffs. Otherwise erroneous results will be
authorized dealers. However, always check the batteries and
recorded.
replace them if necessary prior to making any complaint.
•
Do not press any buttons until the cuff is in position.
•
The appliance should be disposed of according to
•
If the instrument is not used for any length of time, we
Regulation 2002/96/EC-WEEE (Waste Electrical and
recommend removing the batteries.
Electronic Equipment). In case of queries, please con-
tact the municipal authorities responsible for waste disposal
Advice on batteries
in your area.
•
Batteries can be fatal if swallowed. You should therefore
store the batteries and products where they are inaccessible
to small children. If a battery has been swallowed, call a doc-
tor immediately.
•
Batteries should not be charged or reactivated with any other
means, nor should they be taken apart, thrown in the fire or
short-circuited.
•
Remove the batteries from the instrument if they are worn
out or if you are not going to use the instrument for any
length of time. This prevents any damage as a result of leak-
age. Always replace all the batteries at the same time.
14
3. Unit description
Icons in the display:
1. Systolic pressure
1
2
4
3
2. Unit in mmHg
3. Diastolic pressure
4. Pulse symbol
and calculated pulse value
9
WHO
5. Time and date
mmHg
SYS
6. Cardiac arrhythmia symbol
mmHg
DIA
7. Average values of the 3 most recent measurements – AVG
PULSE
/
min
8. Pump up, release air (arrow)
–
+
9. WHO classification
SET
M
+
10. Battery replacement symbol
5
11. Symbol for users 1, 2, 3 and 4
8
7
6
1. Cuff
1
11
2. Cuff line
3. Cuff connector
10
4. Display
9
2
5. START/STOP button
8
6. Memory button M+
3
7. SET button
7
8. Function buttons -/+
6
9. WHO scale
4
5
15
4. Prepare measurement
Proceed as follows to set the date and time:
•
Confirm twice using the SET button.
Inserting battery
•
Use the + or - buttons to first set the month 1–12, and then
•
Remove the battery cover from the
confirm using the SET button.
back of the monitor.
•
Set the day/hour/minute and confirm each using the SET
•
Insert four AA 1.5 V alkaline batter-
button.
ies. Making absolutely sure that you
insert them with the correct polarity
Press the button to exit the settings mode.
as marked. Never use rechargeable
4 x AA (LR06) 1,5
5. Measuring blood pressure
bat teries.
Positioning cuff
•
Replace the battery cover carefully.
Fit the cuff round your bare left
If 4 warning tones have been sounded and the symbol
upper arm. Blood circulation in the
appears simultaneously on the display, it is no longer possible
arm should not be restricted by tight
to perform a measurement and all batteries must be replaced.
clothing or other objects.
Once the batteries have been removed from the device, the
time must be set again. Used batteries should not be disposed
The cuff should be placed on the
of in normal household waste. Dispose of them via your elec-
upper arm so that the lower edge is 2
tronics retailer or your local recycling point.
to 3 cm above the bend of the elbow
You are legally obligated to do so.
and above the artery. The tube should
Note: Batteries containing pollutant substances
be in line with the centre of the palm.
are marked as follows: Pb = Battery contains lead,
Now place the free end of the cuff
Cd = Battery contains cadmium, Hg = Battery
snugly, but not too tightly, around the
contains mercury.
arm, and fix it with the Velcro
Setting date and time
fastener. The cuff should be fitted
tight enough to allow just two fingers
You should set the date and time without fail. Otherwise, you
to fit beneath the cuff.
will not be able to save your measured values correctly with
Insert the cuff tubing into the socket
a date and time and to access them again later. The time is
for the cuff attachment.
shown in 12-hour format i.e. the time from 13:00 is displayed
as 01:00 PM.
16
V
Important: The instrument should only be operated with the
Measuring blood pressure
original cuff. The cuff is suitable for an arm circumference of
•
Put on the cuff as described previously and assume the posi-
22 to 35 cm.
tion in which you want to carry out the measurement.
A larger cuff for upper arm circumferences from 30 to 42 cm is
•
Start the measurement by holding the button . After check-
available from retailers or the service address under order no.
ing the display, during which all numbers light up, the cuff
162.973.
pumps up automatically. Whilst the cuff is pumping up, the
device already calculates measurements for estimating the
Correct posture
necessary pump pressure.
If this pressure is insufficient, the device automatically
pumps more pressure.
•
Then the pressure in the cuff is slowly released and the pulse
is measured.
•
When the measurement has been completed, the remain ing
air pressure is released very quickly. The pulse, the systolic
•
Rest for approx. 5 minutes before each measurement. Other-
and the diastolic blood pressure are displayed.
wise there may be divergences.
•
Measuring can be cancelled at any time by pressing the but-
•
You can perform the measurement either sitting or lying down.
ton .
Always make sure that the cuff is on a level with your heart.
•
If a higher pump pressure is always required, you can bypass
•
To carry out a blood pressure measurement, make sure you
the additional pumping process by holding the M+ button
are sitting comfortably with your arms and back leaning on
shortly after the pumping process starts until the desired cuff
something. Do not cross your legs. Place your feet flat on the
pressure is reached. This should be approx. 30 mmHg above
ground.
the systolic value.
•
In order not to distort the result, it is important to keep still
•
The Err icon appears if it has not been possible to perform
during the measurement and not talk.
the measurement properly. Observe the section in these in-
Select memory
structions on error messages/troubleshooting and repeat the
measurement.
Press the SET button. Select the desired storage space by
•
The device switches off automatically after 3 minutes.
pressing the -/+ function buttons. You have 4 memories with
30 spaces each to store the measurements of 4 different peo-
Wait at least 5 minutes before taking another measurement!
ple separately or to separate measurements taken in the morn-
ing and at night. Confirm your selection with the button
.
17
Evaluating results
ing the measurement. If the icon appears often, please
Cardiac arrhythmia:
contact your doctor. Any self-diagnosis and treatment based
on the test results may be dangerous. It is vital to follow your
This instrument can identify possible cardiac arrhythmia
doctor’s instructions.
disorders during measurement and if necessary indicates the
measurement with the flashing icon
.
WHO classification:
This may be an indicator for arrhythmia. Arrhythmia is a condi-
In accordance with the guidelines/definitions of the World
tion where the heart rhythm is abnormal as a result of defects
Health Organization and the latest findings, the measurements
in the bioelectrical system controlling the heart beat. The
can be classified and assessed according to the following
symptoms (omitted or premature heart beats, slow or exces-
table.
sively fast heart rate) may be caused, among other things, by
However, these standard values serve only as a general guide-
heart disease, age, physical predisposition, excessive use of
line, as the individual blood pressure varies in different people
stimulants, stress or lack of sleep. Arrhythmia can only be as-
and different age groups etc.
certained through examination by your doctor.
It is important to consult your doctor regularly for advice. Your
Repeat the measurement if the flashing icon is displayed
doctor will tell you your individual values for normal blood pres-
after the measurement. Please note that you should rest for
sure as well as the value above which your blood pressure is
5 minutes between measurements and not talk or move dur-
classified as dangerous.
Range of blood pressure values
Systolic (in mmHg)
Diastolic (in mmHg)
Measure
Grade 3: Severe hypertension > = 180 > = 110 Seek medical advice
Grade 2: Moderate hypertension 160 – 179 100 – 109 Seek medical advice
Grade 1: Mild hypertension 140 – 159 90 – 99 Have it checked regularly by doctor
High-normal 130 – 139 85 – 89 Have it checked regularly by doctor
Normal 120 – 129 80 – 84 Check it yourself
Optimal < 120 < 80 Check it yourself
Source: WHO, 1999
The bar graph in the display and the scale on the unit indicate
diastolic pressure in the normal range) the graphic WHO clas-
the range of the blood pressure which has been recorded.
sification on the unit indicates the higher range (high-normal in
If the values for systolic and diastolic pressure are in two dif-
the example described).
ferent WHO ranges (e.g. systolic in the high-normal range and
18
6. Saving, retrieving and deleting results
8. Error messages / troubleshooting
•
The results of each successful measurement are stored
In case of faults, the Err message appears in the display.
together with date and time. With more than 30 items of
Error messages can occur when
measured data, the earliest items of data measured are lost.
1. inflation pressure is higher than 300 mmHg,
•
Use the SET button and then the -/+ buttons to select the
2. blood pressure is unusually high or low,
desired user memory and confirm using the button . Press-
3. you move or talk during the measurement.
ing the M+ button displays the average value AVG of the 3
In the above cases, you must repeat the measurement. Make
most recent measurements stored in the user memory. By
sure that the cuff tube is properly inserted and that you do
pressing the + button, the most recent measurements are
not move or talk. Re-insert the batteries if necessary, or else
displayed; by pressing the - button, the oldest are displayed.
replace them.
•
The average value AVG can only be displayed if there are 3
saved values in the respective user memory.
9. Specifications
•
To clear the memory, you must first select a user memory.
Model no. BM 26
Hold the SET button pressed for approx. 3 seconds. All the
Measurement
Oscillometric, non-invasive blood pres-
values in the current user memory are deleted after 3 brief
method
sure measurement on the upper arm
beeps are output.
Measurement range Cuff pressure 0 – 300 mmHg,
•
If you wish to change the user memory, see “Select memory”.
systolic 60 – 280 mmHg,
7. Cleaning and storing the instrument
diastolic 30 – 200 mmHg,
•
Clean your device and cuff carefully only with a slightly mois-
Pulse 30 – 180 beats/minute
tened cloth.
Display accuracy Systolic ± 3 mmHg, diastolic ± 3 mmHg,
•
Do not use detergents or solvents.
pulse ± 5 % of the value shown
•
On no account must you immerse the computer in water,
Measurement
Max. permissible standard deviation
otherwise liquid can enter it and cause demage.
inaccuracy
according to clinical testing:
•
When storing the device, make sure that no heavy objects
systolic 8 mmHg /diastolic 8 mmHg
are placed on top of it. Remove the batteries. The cuff tube
Memory 4 x 30 memory spaces
should not have any sharp kinks.
Dimensions L 155 mm x W 110 mm x H 70 mm
Weight Approx. 395 g (without batteries)
Cuff size 22 to 35 cm
19
equipment – Part 2 – 30: Particular requirements for the safety
Permissible operating
+10 °C to +40 °C, ≤ 85 % relative air
and essential performance of automated non
-invasive blood
conditions
humidity (non-condensing)
pressure monitors).
Permissible storage
- 20 °C to + 55 °C, ≤ 85 % relative air hu-
•
The accuracy of this blood pressure monitor has been care-
conditions
midity, 800 –1060 hPa ambient pressure
fully checked and developed with regard to a long useful
Power supply
4 x 1,5 V
AA alkaline batteries.
life. If using the device for commercial medical purposes, it
Battery life For approx. 180 measurements,
must be regularly tested for accuracy by appropriate means.
depending on the blood pressure level
Precise instructions for checking accuracy may be requested
and/or pump pressure
from the service address.
Accessories Instruction for use, 4 x 1.5 V AA alkaline
batteries., storage pouch
Classification Internal supply, IPX0, no AP or APG,
continuous operation, type BF applied
part
Technical information is subject to change without notification
to allow for updates.
•
This unit is in line with European Standard EN 60601-1-2 and
is subject to particular precautions with regard to electromag-
netic compatibility (EMC). Please note that portable and mobile
HF communication systems may interfere with this unit. More
details can be requested from the stated Customer Service ad-
dress or found at the end of the instructions for use.
•
This device is in line with the EU Medical Devices Directive
93/42/EC, the “Medizinproduktegesetz” (German Medical De-
vices Act) and the standards EN1060-1 (non-invasive sphyg-
momanometers, Part 1: General requirements), EN1060-3
(non-invasive sphygmomanometers, Part 3: Supplemen-
tary requirements for electro-mechanical blood pressure
measuring systems) and IEC80601-2-30 (Medical electrical
20