CORRECT USE OF THE SLEEVE
1. Find the brachial artery (Arteria Brachialis) of your left arm, which is
located 25 mm above the elbow. Test where your pulse is strongest.
2. Now, place the sleeve around your left upper arm. The hook and
loop adhesive fabric must face outwards and the tube should be
directed towards the hand. The whole arm should be uncovered.
There must not be any pressure on veins from clothing, jewellery,
etc., as this can alter the measurement. Also, do not place the
sleeve over clothing.
3. Place the sleeve so that it is positioned about 2 - 3 cm above the
crease of the elbow. The tube must be in line with the hand, as if the
tube were dividing the hand down the middle. Close the sleeve so
that it surrounds the upper arm firmly. Check that the hook and loop
adhesive fabric is securely fastened.
4. The person whose blood pressure is being taken should now be
sitting on a chair with the left elbow supported on a table.
The tube must not be bent or twisted!
Brachial
Artery
2-3cm
BM122.qxp 9/10/07 2:24 PM Page 18

- 19 -
GB GB
CORRECT BODY POSTURE DURING MEASUREMENT
To obtain as accurate a reading as possible, always remember to hold the sleeve at the same
height as the heart. If the sleeve is higher or lower than the heart this causes deviations from
your actual blood pressure.
Measuring blood pressure while sitting:
• Place your elbow on the table or other surface.
• The sleeve should be at the same height as your heart. If necessary, place a cushion or a
book underneath your lower arm. Keep your arm and the rest of your body in a relaxed
position.
• Any exertion on the body, such as drinking, eating, talking, making or receiving telephone
calls, etc., leads to incorrect measurements and should therefore be avoided.
DESCRIPTION OF THE SYMBOLS IN THE DISPLAY
Symbol in
Condition/ Cause
Corrective measures
the display
Brachial
Artery
ERROR
This symbol appears when an
accurate measurement is
impossible.
This icon appears when the
batteries are almost empty or have
been inserted incorrectly.
Replace all old batteries with
new ones. Always observe the
correct polarity (+ and -) of the
batteries while inserting them.
Repeat the measurement.
Make sure that the sleeve is
attached properly and your
body posture is accurate.
This symbol appears when the
pulse value is detected (blinking)
together with a sound.
BM122.qxp 9/10/07 2:24 PM Page 19

- 20 -
GB GB
INSTRUCTIONS FOR MEASUREMENT
• Inserting the batteries:
Remove the cover of the battery compartment. Observe the
correct polarity while inserting the batteries. At first all
symbols will appear on the LCD display for 3 seconds, then
the date 1 (month) – 1 (day) and the time 12:00.
• Set the date and the time
Press the button
MODE
and the month indicator will flash.
Set the month with the button
SET
.
Press the button
MODE
again and the day indicator will flash.
Set the day with the button
SET
.
Press the button
MODE
again and the hour indicator will flash.
Set the hour (in 12-hour format, for the afternoon and night time a small “P” will appear under
the hour indicator) with the button ı.
Press the button
MODE
again and the minute indicator will flash.
Set the minute with the button
SET
.
Press the button
MODE
once again to return to the time and date indicator.
The device is now ready to operate.
• Steps to measure the blood pressure:
a) Secure the sleeve around your left upper arm (see chapter
Correct Use of the Sleeve
).
b) Sit on a chair and assume an upright body position (see chapter
Correct Body Posture
During Measurement).
c)
Select with the button
SET
the desired number where you would like to save the
measurement. Press the button
START/STOP
to start the appliance. The pumping
process starts automatically. The pressure is raised automatically to 160 mmHg.
If there is a tendency of high blood pressure, the air pressure will automatically be raised
to the required level.
The inflation process can be stopped at any time by pressing
the START/STOP button
.
There should not be more than 2 measurements within a short
period of time as this could alter the measured data.
d) After inflation, the air will escape slowly during the measurement process. The heart
symbol for the pulse will appear on the display.
BM122.qxp 9/10/07 2:24 PM Page 20

- 21 -
GB GB
e) When the blood pressure measurement has been taken, the systolic, diastolic and pulse
values will be shown for 1 minute in the display and the data will automatically disappear
after this time. During the display time, the appliance can be switched off by pressing the
START/STOP
button.
f)
If no button is pressed for 1 minute and no function is used during this time, the energy
saving function will switch the appliance off automatically.
• To save, check and delete the measured data:
a.
Save data:
After each blood pressure measurement, the systolic, diastolic, and pulse data of the
respective measurement will be saved automatically.
The available database can store up to 99 measurements.
If the number of measurements exceeds 99, the oldest measurement will be deleted
automatically.
b.
Check data:
By pressing the
MEMORY
button, the database can be selected and, simultaneously, the
number of full memory spaces will be shown in the upper edge of the LCD display.
If no memory space is taken, NONE is indicated.
Press
MEMORY
button to check the
most recent measurement.
Systolic and diastolic blood pressure, pulse value, time, and date are shown.
Press
MEMORY
button again to check the other measurement values, sorted by save order.
To return to normal display, press
START/STOP
button.
c.
Delete data:
Delete all data:
Remove batteries from the battery compartment and reinsert them. All data will be deleted.
Press
MEMORY
button, to check if the data has been deleted. If done correctly, no data
will be shown in each database.
BM122.qxp 9/10/07 2:24 PM Page 21

- 22 -
GB GB
• Changing the batteries:
If appears in the display or the display is dark, the batteries are dead. Replace the
batteries. Until they are replaced, no further blood pressure measurements are possible.
Change all batteries with battery type R6/LR6/AA.
Only use top quality identical batteries. Do not use new and expired batteries simultaneously. If
the batteries have leaked, remove them from the battery compartment using a cloth and
dispose them according to regulations. Wipe battery compartment and other compartments
carefully with a cloth. Be sure not to come in direct contact with the battery acid..
Batteries do not belong in your household waste.
MAINTENANCE
• To clean the casing, use a dry cloth moistened with water or a mild cleaning agent, and wipe
the unit using a dry cloth. If the sleeve is soiled and needs to be cleaned, use a dry or slightly
damp cloth only.
• Do not use any spray cleansers or strong cleaning agents.
• If the unit is not going to be used for an extended period of time, remove the batteries
(expired batteries can damage the unit).
STORAGE
• Always store the appliance in its tray after use.
• Protect the measuring unit and its accessories from direct sunlight, high temperatures,
humidity, and dust.
• Do not subject the unit to extremely low (below -20°C) or high (over 50 °C) temperatures.
• Do not fold the sleeve.
PREVENTION OF MALFUNCTIONING
• Do not twist the sleeve.
• Do not pump air into the sleeve unless it is wound around the arm.
• Do not try to disassemble the unit or to change the components of the appliance or sleeve.
• Do not drop the measuring unit or allow it to suffer any heavy impact.
BM122.qxp 9/10/07 2:24 PM Page 22

- 23 -
GB GB
TAKE NOTE
•
Caution:
The sleeve size only covers a size range of 24 to 32 cm in diameter and is only to
be used by adults.
• Do not use the measuring unit if your arm is injured.
• If the measuring appliance should continue inflating, open the sleeve immediately or press
the
START/STOP
button.
• By pressing
START/STOP
during inflation, the inflation process is stopped and the air is
immediately released from the sleeve.
• Should by any means pain occur during the measurement, stop immediately by pushing the
START/STOP
button or by opening the sleeve.
TECHNICAL DATA
Measuring procedure
Oscillometrical
Measurement range
Pressure 0 ~ 300 mmHg, Pulse 40 ~ 199 beats per minute
Measurement accuracy
Pressure +/- 3mmHg, Pulse max. +/- 5%
Inflation
Electric rolling pump
Deflation Magnetic
valve
Display
Systole, Diastole, pulse value, year/date/time, memory database
Memory Database
with 99 memory spaces
Automatic switch off
1 minute after last pressing of any button
Weight
Approx. 265g with batteries
Sleeve size
Arm diameter of about 24 to 32 cm
Operating temperature range +10°C to +40°C, <85% relative humidity
Storage temperature range
-20°C to +70°C, <85% relative humidity
Power supply
6V DC 4 x 1,5 V R6/LR6/AA alkaline batteries
Battery life
300 measurements (2 per day)
Measures
80 x 120 x 40 mm
BM122.qxp 9/10/07 2:24 PM Page 23

- 24 -
GB GB
INSTRUCTIONS ON ENVIRONMENT PROTECTION
Do not dispose of this product in the usual household garbage at the end of its life
cycle; hand it over at a collection point for the recycling of electrical and electronic
appliances. The symbol on the product, the instructions for use or the packing will
inform about the methods for disposal.
The materials are recyclable as mentioned in its marking. By recycling, material recycling or other
forms of re-utilization of old appliances, you are making an important contribution to protect our
environment.
Please inquire at the community administration for the authorized disposal location.
GUARANTEE AND CUSTOMER SERVICE
Before delivery our devices are subjected to rigorous quality control. If, despite all care, damage
has occurred during production or transportation, please return the device to your dealer. In
addition to statutory legal rights, the purchaser has an option to claim under the terms of the
following guarantee:
For the purchased device we provide 2 years guarantee, commencing from the day of sale. During
this period we will remedy all defects free of charge, which can be demonstrably attributed to
material or manufacturing defects, by repair or exchange.
Defects which arise due to improper handling of the device and malfunctions due to interventions
and repairs by third parties or the fitting of non-original parts, are not covered by this guarantee.
BM122/0708
ELTA GmbH
Carl-Zeiss-Str. 8
D- 63322 Rödermark/
Ober-Roden, Germany
0197
BM122.qxp 9/10/07 2:24 PM Page 24

- 25 -
GB GB
NOTE
This blood pressure measuring device conforms to European regulations and carries the CE-
Symbol “CE 0197”.
The quality of this device has been proven and conforms to the requirements of the European
Community regulation 93/42/EEC (EN IEC 60601-1 – general safety requirements, EN IEC
60601-1-2:2001 – electromagnetic compatibility; requirements and screening processes) from
July 14, 1993 according to medical appliance’s and service standards as follows:
EN 1060-1:
Non-invasive blood pressure device – general requirements
EN 1060-3:
Non-invasive blood pressure device – additional requirements for electromagnetic
blood pressure appliances.
EN 1060-4:
Non-invasive blood pressure device – screening process to determine the
measurement accuracy of automatic non-invasive blood pressure appliances.
Classifications:
• Device with internal power supply
• BF-type device
• IPXO
• Device is not suitable for use near inflammable anaesthetics mixtures with air or
oxygen or nitrogen.
• Continuous operation with short term charging