Labconco VACUUBRAND Diaphragm Pump 7393001: Troubleshooting

Troubleshooting: Labconco VACUUBRAND Diaphragm Pump 7393001

background image

page 50 of 74

Fault

Possible cause

Remedy

❑ 

Pump does not 

start or stops im-

mediately.

Electrical power cord 

not plugged in, electri-

cal supply failure?

Plug in power cord. 

Check fuse.

Device fuse blown?

Identify cause of failure. 

Replace device fuse.

Overpressure in outlet 

line?

Remove blockage in line, 

open valve.

Motor overloaded?

Allow motor to cool down, 

identify and eliminate 

cause of failure. Manual 

reset is necessary. Switch 

off pump or unplug.

Pump does not 

achieve its ultimate 

vacuum or usual 

pumping speed.

Centring ring at small 

flange connection not 

correctly positioned, or 

leak in the pipeline or 

vacuum system?

Check pump directly - 

connect vacuum gauge 

directly at pump inlet - 

then check connection, 

pipeline and vacuum 

system if necessary.

Long, narrow vacuum 

line?

Use lines with larger di-

ameter, length as short as 

possible.

Pump has been ex-

posed to condensate?

Allow pump to run for 

some minutes with atmo-

spheric pressure at the 

inlet to purge.

Deposits have been 

formed inside the 

pump?

Clean and inspect the 

pump heads.

Diaphragms or valves 

damaged?

Replace diaphragms and/

or valves.

Outgassing substances 

or vapor generated in 

the process?

Check process parame-

ters.

Troubleshooting

background image

 page 51 of 74

Fault

Possible cause

Remedy

Pump too noisy.

Atmospheric or high 

pressure at the pump 

inlet?

Connect hose or silencer 

to pump outlet. Be careful 

not to cause outlet over-

pressure, especially with 

condensable vapors.

Diaphragm crack or 

diaphragm clamping 

disc loose?

Perform maintenance.

Other than above men-

tioned causes?

Contact local distributor.

Pump seized.

Contact local distributor.

 A service manual with exploded view drawings, spare parts list and 

directions for repair is available on request.

The service manual is intended for trained service people only.