Labconco VACUUBRAND Diaphragm Pump 7393000: Troubleshooting

Troubleshooting: Labconco VACUUBRAND Diaphragm Pump 7393000

page 50 of 74

Troubleshooting

Fault Possible cause Remedy

Pump does not

Electrical power cord

Plug in power cord.

start or stops im-

not plugged in, electri-

Check fuse.

mediately.

cal supply failure?

Device fuse blown?

Identify cause of failure.

Replace device fuse.

Overpressure in outlet

Remove blockage in line,

line?

open valve.

Motor overloaded?

Allow motor to cool down,

identify and eliminate

cause of failure. Manual

reset is necessary. Switch

off pump or unplug.

Pump does not

Centring ring at small

Check pump directly -

achieve its ultimate

ange connection not

connect vacuum gauge

vacuum or usual

correctly positioned, or

directly at pump inlet -

pumping speed.

leak in the pipeline or

then check connection,

vacuum system?

pipeline and vacuum

system if necessary.

Long, narrow vacuum

Use lines with larger di-

line?

ameter, length as short as

possible.

Pump has been ex-

Allow pump to run for

posed to condensate?

some minutes with atmo-

spheric pressure at the

inlet to purge.

Deposits have been

Clean and inspect the

formed inside the

pump heads.

pump?

Diaphragms or valves

Replace diaphragms and/

damaged?

or valves.

Outgassing substances

Check process parame-

or vapor generated in

ters.

the process?

page 51 of 74

Fault Possible cause Remedy

Pump too noisy.

Atmospheric or high

Connect hose or silencer

pressure at the pump

to pump outlet. Be careful

inlet?

not to cause outlet over-

pressure, especially with

condensable vapors.

Diaphragm crack or

Perform maintenance.

diaphragm clamping

disc loose?

Other than above men-

Contact local distributor.

tioned causes?

Pump seized.

Contact local distributor.

A service manual with exploded view drawings, spare parts list and

directions for repair is available on request.

+ The service manual is intended for trained service people only.