Alexika RET control-visc – page 2

Alexika
RET control-visc

Manual for Alexika RET control-visc

background image

21

>100 mm

>100 mm

>100 mm

Safety instructions

•  

Please read the instruction manual in full before use and 

follow the safety instructions.

•  Keep the instruction manual in a place where it can be accessed 

easily.

General instructions

•  

Ensure that only trained staff use the device.

• 

 Be sure to comply with all safety instructions, directives and all 

matters of health, safety and accident prevention in the workplace.

•  The plug must be earthed (grounded contact).

  

Caution - magnetism!

  

Beware of possible effects from the 

magnetic field (pacemakers, data media, 

etc.).

   

  

Burns hazard!

Exercise caution when touching housing  

components and the heating plate. The 

heating plate can reach temperatures of up to 320°C. Watch for 

residual heat after appliance has been switched off.

   

The device should only be transported once it has cooled!

Device configuration

  

Do 

not

 operate the device in explosive 

atmospheres, in the presence of hazard-

ous materials or under water.

•  Install the device on a flat, stable, clean, non-slip, dry and fire-

resistant surface.

•  The device feet must be clean and undamaged.

•  The power supply cable and cables to the external sensors must 

not be allowed to come into contact with the heated mounting 

plate.

•  Prior to each use, always check the device and accessories for 

damage. Never use damaged parts.

Permissible media / Contamination / Side reactions

   

  

Caution!

 Only media whose flashpoint 

lies above the set safety temperature 

limit (50°C  - 380°C) may be processed 

or heated with this device.

The set safety temperature limit must be at least 25°C below the 

flashpoint of the medium in use.

   

  

Pay attention to the risks associated 

with:

  -  flammable materials

  -  flammable media with low steam pressure

  -  glass breakage

  -  incorrect vessel size

  -  overfilling of medium

  -  unstable vessel.

•  Only process pathogenic materials in closed vessels under a suit

-

able extractor hood. Please contact 

IKA

®

 if you have any ques-

tions.

Only work with media for which the en-

ergy  input during processing is harmless. 

This also needs to take into account 

other sources of energy such as, for example, light irradiation.

•  At high speeds the mounting plate is liable to warming, even if 

the device is not in heating mode. 

•  Bear in mind the possibility of contamination or unwanted chem

-

ical reactions.

•  There is the possibility that friction from rotating accessory parts 

can result in contamination of the medium being processed.

•  Bear  in  mind  the  following  when  using  PTFE-coated  magnetic 

rods: 

Chemical reactions of PTFE occur in contact with molten 

or solute alkali metals and alkaline earth metals as well as with 

fine powders of metals in groups 2 and 3 of the periodic table 

ATTENTION

DANGER

at temperatures above 300-400°C. Only elementary fluorine, 

chlorine trifluoride and alkali metals attack it; halogenated hy-

drocarbons have a reversible swelling effect

.

  

  

(Source: Römpp “Chemie-Lexikon” and “Ullmann” Vol. 19)

•  Only glass-coated magnetic rods should be used in conjunction 

with solute alkali metals or alkaline earth metals or at tempera-

tures above 250°C.

Performing trials

  Always

 wear personal protective 

  equipment in accordance with the hazard  

  class of the media being worked with. 

Otherwise there is the risk of the following:

  -  spraying and evaporation of liquids 

  -  ejection of parts

  -  release of toxic or inflammable gases.

•  If any of the following happens, reduce the speed:

  -  medium sprays out of vessel due to excessive speed

  -  process is not smooth

  -  vessel moves around on the mounting plate.

Accessories

•  Safe operation can only be ensured when working with acces

-

sories as described in the “

Accessories

” section.

•  Only put together accessories with the power supply cable dis

-

connected. 

•  Refer to the operating instructions for the accessories.

•  Immerse the external temperature sensor (PT1000, ETS-D, etc.) 

at least 20 mm into the medium.

•  The PT1000 external temperature sensor must always be inside 

the medium.

•  Accessories must be connected securely to the device and must 

not come loose on their own. The centre of gravity of the device 

must be located within the mounting plate.

Power supply / switching off the device

•  The specified settings on the rating plate must coincide with the 

actual power supply.

•  The plug for the power supply connection must be easy to access 

and remove.

•  The device is only disconnected from the power supply when the 

power or device plug is removed.

  

After an interruption to the power sup-

ply, the device starts up again in Mode B.

Protection of the device

•  The device must only be opened by trained, skilled personnel.

•  Do not cover the device, even partially, with elements such as 

metallic plates or sheets otherwise it may overheat.

•  Avoid knocks or impact to the device or accessories.

•  Ensure the mounting plate remains clean.

•  

Observe minimum distances between devices, between the device 

and the wall, and above and below the device (min. 800 mm).

WARNING

DANGER

WARNING

WARNING

DANGER

WARNING

background image

22

Unpacking

• Unpacking

  -  Unpack the device carefully

  -  Any damage should be notified immediately to the shipping 

agent (post office, railway network or transport company)

• Scope of delivery

-  RET control visc heated magnetic stirrer

  -  Power supply cable

  -  Operating instructions

  -  Screwdriver (safety circuit)

  -  USB cable

  -  Protective cover: 

H104

  -  Magnetic rods: 

IKAFLON

30

 and 

40

mm

  -  Temperature sensor: 

PT 100.70

 Correct use

 Assembling the support rod

Note: 

When using bath attachments with a diameter 

over 180 mm, use sup-

port rod H 16 V in con-

junction with an  H 

16.1 extension.

The set safety temperature is displayed permanently (

7

). 

Factory settings: approx. 380°C 

Range: 50 - 380°C 

Setting the safety temperature limit

a

   

Set the main switch 

(A)

 to the ON position

a

   

Set the temperature limit using the screwdriver (supplied) 

at point 

(B)

The set value is displayed on the screen

The adjustable maximum heating plate temperature of 340°C is 

at least 10°C under the set safety temperature limit. 

Caution: the set safety temperature limit must always be 

at least 25°C below the flashpoint of the medium in use.

Functional test: safety circuit switch-off

a

   

Heat the device to a temperature of over 50°C.

a

   

Set the safety temperature limit as far to the left as possible (50°C) 

and set the main switch (A) to the ON position

The display will show:        Err 25

 Setting the safety temperature limit

Be careful not to turn the adjustment screw too far to the left or 

right or this will damage the potentiometer.

• Use

   -  for mixing and/or heating liquids.

• Area of use (only indoors)

   -  laboratories  

- schools

   -  pharmacies  

- universities

The device is suitable for use in all areas apart from domestic areas 

and areas directly connected to a low-voltage supply which also 

serves domestic areas.

The safety of the user cannot be guaranteed if the device is used 

in conjunction with accessories which are not supplied or recom-

mended by the manufacturer, if the device is used contrary to the 

intended purpose against the manufacturer’s instructions or if the 

device itself or PCB has been subjected to modifications by third 

parties.

•  

Remove the threaded plug 

(P)

•  Remove the protective cap from the support rod

•  Place the washer between the housing and the nut

•  Screw in the support rod manually until it cannot be 

tightened further

•  Tighten the nut using an SW17 open-end spanner

•  Assemble the accessories using a boss head clamp

background image

23

Remove the protective film from the mounting plate before use! The “Back” menu key is used to change between menu items.

Start-up

Stirring

Heating

Connecting an ex-

ternal thermometer

(directtemperature 

control in the me-

dium)

Weighing

pH measurement

Timer

Set the main switch (A) to the OFF position

Insert the power supply cable into the mains socket (K)

Set the main switch (A) to the ON position

Standard mode A is selected (see “Operating modes”)

Set the speed using the rotary knob (D)

The set value is shown on the display (F)

Press the rotary knob (D) to start the stirring function

i

The symbol (16) appears when the motor is active

  F 

Set the safety temperature limit (see “Setting the safety temperature limit”)

  F 

Set the target temperature using the rotary knob (C)

  a 

The set value is shown on the display (J)

  F 

Set the temperature control mode (see “Setting the temperature control mode”)

  F 

Press the rotary knob (C) to start the heating function

    

The target and current temperature are shown permanently on the display (J):

i

When the heating function is switched on, the heating symbol (4) is displayed

    

The maximum adjustable heating plate temperature setting is 340°C

i

Whenever the mounting plate is at a temperature above 50°C, the display (J) shows                   and the current tem-

perature is shown, even if the device is switched off.

Set the main switch (A) to the OFF position

Connect the safety contact thermometer compliant with DIN 12878 Class 2 or temperature sensor PT 100 (individual 

sensor) or PT1000 (double sensor) to socket (L)

Set the main switch (A) to the ON position

i

   Temperature sensor PT 100 

a  

The current temperature (5) shown on the display (J) corresponds to the tempera-

ture of the medium. The “Probe” symbol (11) lights up.

i  

Temperature sensor PT 1000 

a  

The current temperature (5) for the temperature sensor shown on the display (J) 

corresponds to the temperature of the medium. The “External temperature probe” 

symbol (11) appears. The temperature of the carrier fluid (3) is also displayed.

Press the (G) key

Wait until the system is stable (once the system is stable, a “g” appears after the digit 0).

Place the weight in the centre of the heating plate and wait until the system is stable once again.

a

The maximum display range is up to 2,000 g 

a

The weighing range is max. 5,000 g 

a

It is only possible to use the weighing mode when the heating or stirring functions are not active

a

The weighing module can be calibrated (see section entitled “Calibrating the weighing module”)

a

see section entitled “Calibration of the pH probe and pH measurement”

Press the (H) key 

Set the time in hours, minutes and seconds by turning the knob (D) and confirm by pressing it; the timer function 

starts to count down the remaining time once the heating or stirring function has started (see “Timer function”)

i  

Once the timer reaches 0, the heating function is switched off, even if the motor has only just been started.

    

If the heating is activated during the timer sequence, only the heating is switched off, but the motor continues 

to run (delay in boiling). 

Operation 

(brief instructions)

 Setting the operating mode

Operating the device in A, B, or C 

Factory setting: Mode A

The set mode is shown permanently on the display (10).

Mode A 

All the settings are retained when the device is switched off or 

disconnected from the mains. When the device is switched on, 

the status of the stirring and heating functions is set to (OFF). 

Mode B 

All the settings are retained when the device is switched off or 

disconnected from the mains. When the device is switched on, 

the status of the stirring and heating functions is set to whatever 

it was before the device was last switched off (ON or OFF).

background image

24

Mode C

In this mode, the device can be used either with or without a 

temperature sensor.

When the device is switched on using the main switch, the 

stirring and heating functions are started in the last set status 

and at the last set values. The set nominal values in operating 

modes A and B cannot be modified.

Changing the operating mode

-  

Scroll down to the “Mode” icon in the menu

-  

Open the submenu by pressing the rotary knob (D)

-  

Use the rotary knob (D) to select the desired mode and 

confirm by pressing the rotary knob (D)

-  

A short description of the operating mode is then shown 

on the display

-  

Confirm selection of the operating mode by pressing the 

rotary knob (D) once more.

 Functions

When using the PT100, the selected temperature control mode is 

shown on the digital display by the 

 (11) symbol.

The target temperature (2) and the 

current temperature (5) shown on 

the display refer to the temperature 

of the medium. The heating plate 

temperature is limited by the set 

safety temperature (7). The medium 

temperature is controlled by means 

of a microcontroller. The medium 

temperature is measured by the PT100 

temperature sensor and is heated 

up as quickly as possible without 

overshooting to the set temperature. 

The microcontroller is capable of 

adjusting automatically to the different 

heating capacities of various vessels 

and their contents. As a result, the heat 

is distributed in an optimum fashion 

with no temperature drift or ripples.

Regulation of the temperature of the medium 

using PT100 temperature sensor

The user is able to set the device to three different methods 

depending on their specific needs. The selected temperature 

control method is displayed alongside the symbol (

11

).

Control methods

Changing between the 

control methods

-  

Scroll down to the 

“Heating” icon in the menu

-  

Open the control method 

submenu by pressing the 

rotary knob (D)

-  

Use the rotary knob (D) to 

select the desired mode 

and confirm by pressing 

the rotary knob (D)

-  

A short description of the 

operating mode is then 

shown on the display

-  

Confirm selection by 

pressing the rotary knob (D) 

once more.

PID

:  For use with very sensitive materials which need to be 

handled carefully.

  Advantages:

  •    very accurate warming 

of the liquid

  •    no exceeding the target 

temperature

  Disadvantages:

  •    slow heating operating

T

Set

t

Fast PID:  

combines the advantages of the PID and two-point 

methods (very fast 

heating and high 

accuracy; possible 

overheating).

Two-point

:

     For media which is not too sensitive.

  Advantages:

  •    Heating much more 

rapid than 

PID

  Disadvantages:

  •    Possible overheating of 

up to 10 K

T

Set

t

T

Set

t

background image

25

Operation with the PT1000 double 

sensor is indicated on the digital 

display by the 

 symbol. One 

temperature sensor is located in the 

heat carrier fluid and the other is 

located in the reaction medium. The 

set target temperature (2) refers to 

the temperature of the medium. The 

higher temperature is automatically 

attributed to the heat carrier fluid and 

shown on the display as an additional 

current temperature with the symbol 

 (3).

The actual temperature of the reaction 

medium is shown as the current 

temperature (5). The heating plate 

temperature (and therefore the heat 

carrier fluid temperature) are limited by the set heating plate 

safety temperature limit (7). The reaction medium temperature 

is limited by the set temperature limits (carrier fluid/reaction 

temperature limit).

    

see section entitled “Limits”

Limits

In order to ensure that neither the reaction fluid nor the carrier 

fluid overheat, the user can set separate temperature limits.

•  Scroll to the “Heating” icon 

in the menu

•  Open the submenu by 

pressing the rotary knob (D)

•  Open the “Limits” submenu 

followed by “Medium 

limit” and then set the 

temperature limit for the 

reaction fluid.

•  Confirm the set limit by 

pressing the rotary knob (D)

•  Open the “Carrier limit” 

submenu and set the limit 

temperature for the carrier 

fluid.

•  Confirm the set limit by 

pressing the rotary knob (D)

Regulation of the temperature of the medium 

using PT1000 double temperature sensor

•  Open the “Limit 

information” submenu. 

All set limit temperatures, 

including the safety limit 

temperature of the heating 

plate, as displayed on one 

screen.

•  Close the submenu by 

pressing the rotary knob (D)

Calibrating the temperature sensor

In order to use tolerances to limit temperature deviations, the 

user can calibrate the temperature sensor in conjunction with the 

device.

2-point calibration:

Calibration using two temperatures

3-point calibration:

Most accurate way of calibrating the temperature sensor at three 

different temperatures

Note:

A calibration carried out by the user is indicated on the display 

with the symbol (11) shown in 

red

.The calibration of the PT100 

and PT1000.50 double 

sensor are carried out 

separately.

•  Scroll to the “Heating” icon 

in the menu

•  Open the submenu by 

pressing the rotary knob (D)

•  Open the “Temperature 

sensor” submenu and 

then “Calibration”, finally 

confirming by pressing the 

rotary knob (D)

•  Select either 2-point 

calibration or 3-point 

calibration and confirm by 

pressing the rotary knob (D)

In this function, the speed can be pre-set before starting the 

calibration.

background image

26

g

g

Stirring bar breakaway

If the device detects that a stirring bar has broken away, the 

stirring function is interrupted temporarily in order to pick up 

the bar. The system then returns to the previously set speed. 

When the system detects a bar breakaway, this is indicated with 

a special symbol (

 instead of symbol 16). If another stirring 

bar breakaway occurs within the next 3 minutes (as long as the 

target speed has not been modified), once the bar is recovered 

the system only goes up to a speed of 100 rpm. In the case of 

4 consecutive stirring bar breakaways -with less than 3 minutes’ 

normal speed between them - the error message ER 41 is 

displayed. The speed is reduced to 200 rpm and the heating is 

switched off.

In such cases, the testing conditions will need to be reconsidered.

NOTE:

The user is responsible for ensuring the suitability/function 

of the stirring bar breakaway monitoring for the current 

conditions: stirring bar, speed, vessel, medium, etc. 

Stirring bar breakaway monitoring does not work if the 

stirring bar already breaks away before the system has 

reached the set target speed!

The higher the torque to be transferred, the more safely the 

stirring bar breakaway monitoring works.

In a glass beaker with water, the stirring bar breakaway for a 

bar from 40 mm (d=8 mm) upwards can be seen from 300 - 

1,200 rpm. Longer stirring bars, larger bar surface areas, e.g. 

Trika, or higher speeds increase the torque to be transferred.

•  Scroll to the “Stirring” icon 

in the menu

•  Open the submenu by 

pressing the rotary knob (D)

•  Open the “Stirring bar 

breakaway” submenu and 

confirm by pressing the 

rotary knob (D)

•  Acknowledge the informa

-

tion screen displayed by 

pressing the rotary knob (D)

g

Intermittent mode

Intermittent mode consists of regular, cyclical interruption of 

the stirring function. The run time and interval can both be set. 

When switching on the device, the time for each is displayed if 

intermittent mode is activated

•  Scroll to the “Stirring” icon 

in the menu

•  Open the submenu by 

pressing the rotary knob (D)

•  Open the “Interval” 

submenu and confirm by 

pressing the rotary knob (D)

•  Enter the run time and stop 

time using the rotary knob 

(D) and finally confirm the 

two by pressing it again.

Torque trend measurement

The torque trend measurement is used to deduce the change in 

viscosity of the reaction medium. 

The devices are 

not

 designed to measure absolute viscosity. They 

only measure and display the relative change in the viscosity of the 

medium from a starting point specified by the user.

Once the speed of the motor and magnetic rod in the medium 

have stabilised to the target speed, the viscosity measurement is 

started at 100%.

Consequently, an increase in the measured torque trend is indicated 

by a value > 100% and a reduction in values < 100%.

The value can always be reset to 100% by pressing the “Back” (F) 

membrane key.

Note:

Torque trend measurement only works for a constant set 

speed for the duration of the measurement.

As a result, intermittent mode cannot be used in 

conjunction with torque trend measurement.

background image

27

The current control variable is saved as the reference 100% ΔP 

and shown on the digital display. The change in the viscosity is 

then shown in %. Depending on whether the viscosity increases 

or decreases, the percentage rises or falls above or below 100% 

respectively.

•  Scroll to the “Stirring” icon 

in the menu

•  Open the submenu by 

pressing the rotary knob 

(D)

•  Open the “Torque trend 

measurement” submenu 

and confirm by pressing 

the rotary knob (D)

•  Acknowledge the informa

-

tion screen displayed

Weighing module

The weighing module is a function which allows the user to 

perform simple weighing 

tasks.

•  

To start the weighing pro-

cess, press the (G) button 

on the front membrane or 

alternatively select Weig-

hing from the menu. Press 

the rotary knob (D) to start 

the weighing function.

Note:

The heating and stirring 

functions must be deacti-

vated.

•  

It may take several 

moments for the system to 

stabilise.

  

Once the system is stable, 

a “g” will be shown 

alongside the measured 

value.

Weighing - Calibration

•  Open the “Calibration” 

submenu and confirm by 

pressing the rotary knob 

(D)

•  Enter the calibration 

weight and confirm by 

pressing the rotary knob 

(D)

•  Place the calibration 

weight on the mounting 

plate and wait until the 

device indicates that the 

calibration process has 

been completed. 

  

This process needs to be 

repeated two or three 

times before closing cali-

bration mode 

  

Once the calibration 

process has been 

completed successfully, the 

weighing module is ready 

for use.

  

Regularly re-calibrate the 

device.

Timer / Counter

The Timer function allows the user to stop the heating process 

after a certain time has lapsed. The Timer function can be set for 

up to 99 hours, 59 minutes and 59 seconds. 

Note:

The time starts to count down when starting both functions 

(heating and/or stirring).

Once the heating function is activated, the timer function only 

refers to heating. The stirring function continues to operate as 

previously, thereby avoiding a delay in boiling in the liquid. The 

motor is only switched off when the heating function is not 

active. The stirring function is stopped if the heating function is 

not started during the timer cycle and the time runs out. 

NOTE:

By entering a value of 

00:00:00

 (hh:mm:ss), the counter is 

activated. The counter shows the time elapsed during the current 

experiment.

background image

28

Setting of the timer

•  

To start the timer function, 

press the (H) button on 

the front membrane or 

alternatively select “Timer” 

from the submenu.

  

The timer settings are 

opened by clicking on 

“Set”. The timer can be set 

by turning the rotary knob 

(D).

•  Set the hours and press the 

rotary knob (D)

•  Set the minutes and press 

the rotary knob (D)

•  Set the seconds and press 

the rotary knob (D)

  

The selected timer (count-

down) time is shown on 

the display.

Safety settings

The user has several options 

in the safety settings menu, 

and the device functions can 

be set according to the user’s 

individual requirements.

•  

To start, select “Safety” in 

the submenu and confirm 

by pressing the rotary knob 

(D).

  

The “Timer” function is used to 

define how long an interruption 

in communications between the 

magnetic stirrer and the wireless 

controller box can be before there 

is a reaction in the magnetic stirrer.

  

If the set time limit is exceeded, 

the pre-selected temperature and 

speed values become active.

  

Please refer to the separate in-

struction manual for the wireless 

controller box for more details.

  

In the other menu items, the 

user can enter limit values for the 

temperature and speed.

•  

Select “Password” from the 

submenu and confirm by 

pressing the rotary knob 

(D). 

Enter a combination of 

digits as your password and 

confirm by pressing the 

rotary knob (D).

  

This information screen 

informs the user that they 

must confirm the hot 

plate safe temperature 

value when starting up the 

device.

•  Safety circuit confirmation 

at device start-up: the hot 

plate safe temperature 

value must be confirmed by pressing the rotary knob (D).

pH measurement 

The device has an integrated pH measuring feature for 

determining the pH of different liquids.

The pH sensor must be calibrated before being used to attempt a 

pH measurement.

The calibration is used to adjust the pH probe and the device 

so that they work together correctly. As part of the process, 

the neutral and pH gradient are specified for the measurement 

chain. To complete the calibration, use pH

4, pH5 and pH7 buffer 

solutions in accordance with DIN 19266.

Note:

a pH calibration can only be carried out with an inserted 

temperature sensor.

•  Scroll

 down to the “pH 

probe” icon in the menu

•  Open the submenu by 

pressing the rotary knob (D)

•  Open the “Calibration” 

submenu and confirm by 

pressing the rotary knob (D)

•  Place the pH probe and 

temperature sensor in the 

required buffer solution.

•  Once the value is stable, 

confirm it by pressing the 

rotary knob (D).

•  Follow the instructions on 

the display and place the 

pH probe and temperature 

sensor in the next buffer 

solution.

background image

29

  

Confirm once more the 

value displayed.

•  In the case of a 3-point 

calibration, a third buffer 

solution must be used.

  

pH measurements can be 

carried out once the cali-

bration has been comple-

ted successfully.

  

Please note:

  

the use of the temperature 

sensor is limited by the 

maximum permissible tem-

perature.

  

Ensure that the 

temperature of the liquid does not exceed the maximum 

permissible temperature.

Settings

The “Settings” menu can 

be used to change how 

the device operates in 

accordance with the user’s 

requirements. 

 Connections and interfaces

Micro-USB

Bluetooth

Bluetooth 

The device allows the user to establish a connection to a wireless 

controller box. 

The device can then be 

operated via a wireless 

controller box within a 

range of 10 m (without 

obstructions).

Establishing a connection 

Please refer to the separate instruction manual for the wireless 

controller box for more details.

background image

30

Power supply 

connection

PT 100 / PT 1000 plug

RS 232

BNC connector 

(only for pH probe)

Pt1000-2

5

3

1

4

2

Pt1000-1

Pt100

5

3

1

4

2

9-pin connector

Serial RS 232 interface and USB 

Configuration

The RS 232 serial interface can be used to operate the device 

using a computer and a suitable user program (labworldsoft).

In order to increase safety when operating the RET using a 

PC, a watchdog function can be activated which monitors 

the continuous data flow (see section entitled: “Watchdog 

function”).

•   The functions of the interface circuit between the laboratory 

device and the automation system are a selection from the 

signals specified in the EIA standard RS232 as per DIN 66020 

Part 1. The assignment of the different signals can be seen in 

the image.

•   Standard RS 232, corresponding to DIN 66259 Part 1 is valid 

for the electronic characteristics of the interface circuits and 

assignment of signal states.

•   Transmission process: asynchronous character transmission in 

start-stop operation.

•   Transmission type: full duplex

•   Character format: character composition according to data 

format in DIN 66022 for start-stop operation. 1 start bit, 

7 character bits, 1 parity bit (even), 1 stop bit.

•   Transmission speed: 9,600 bit/s

•   Data flow control: hardware handshake RTS/CTS

    

RTS: (PIN 7) LOW (positive voltage)

   

/ COMPUTER can transmit

    

RTS: (PIN 7) HIGH (negative voltage)

   

/ COMPUTER cannot transmit

    

CTS: (PIN 8) LOW (positive voltage)

   

/ COMPUTER ready-to-receive

    

CTS: (PIN 8) HIGH (negative voltage)

   

/ COMPUTER not ready-to-receive

•  Access method: data transmission from the device to the 

computer only occurs after a request from the computer.

Command syntax

The following points should be noted for the instruction set:

•   Commands are generally sent from the computer (master) to 

the lab device (slave).

•   The lab device only responds to requests from the computer.

•   Commands and parameters, as well as consecutive parameters, 

must be separated by at least o n e space (code: hex 0x20).

•   Each individual command including parameters and data and 

all responses are completed with CR LF (code: hex 0x0D and 

0x0A) and can have a maximum length of 10 characters.

•   The dot is used for decimal separators in a floating-point value 

(code: hex 0x2E).

The details given above generally comply with the recommendations 

of NAMUR (NAMUR recommendations for the design of electrical 

plug-in connectors for analogue and digital signals in laboratory 

MSR devices. Rev. 1.1).

Summary of available NAMUR commands

Abbreviations used:

X,y =  numbering parameter (integer)

m  =  variable value, integer

n  =  variable value, floating-point number

X  =  1 medium temperature(external temperature sensor)

X  =  2 hot plate  temperature

X  =  3 hot plate safety temperature

X  =  4 speed

X  =  5 viscosity trend

X  =  7 heat transfer medium temperature

X  =  80 pH value 

X  =  90 Weight value

X  =  54 Error 5 response time in seconds (180 <= n <= 1200)

X  =  55 Intermittent mode cycle time in seconds 

       ( 10 <= n <= 600 ) 

X  =  56 Intermittent mode, duration of interruption in seconds 

       ( 5 <= n <= 60 )

background image

31

NAMUR commands

Function

Display 

(additional)

IN_NAME

Title request

IN_PV_X  X=1;2;3;4;5;7;80;90 Current value reading

IN_SOFTWARE

Software ID number, date, version request

IN_SP_X  

X=1;2;3;4;7;8;12; 

42;54;55;56;

Set target value reading

IN_TYPE

Lab device identification request

OUT_NAME name

Output of identification name. (Max. 6 characters; default: IKAMAG)

OUT_SP_12@n

Setting WD safety limit temperature with set value echo

OUT_SP_42@n

Setting WD safety limit speed with set value echo

OUT_SP_X n  

X=1;2;4;7; 

    54;55;56

Setting of target value to n

OUT_WD1@m

Watchdog mode 1: if event WD1 should occur, the heating and stirring functions are 

switched off and ER 2 is displayed. Set watchdog time to m (20 - 1,500) seconds, 

with watchdog time echo. This command launches the watchdog function and must 

be transmitted within the set watchdog time.

OUT_WD2@m

Watchdog mode 2: if event WD2 should occur, the speed target value is changed to 

the WD safety speed limit and the temperature target value is changed to the WD 

safety temperature limit value. The warning WD is displayed. The WD2 event can be 

reset with OUT_WD2@0 - this also stops the watchdog function. Set watchdog time 

to m (20 - 1,500) seconds, with watchdog time echo. This command launches the 

watchdog function and must be transmitted within the set watchdog time.

RESET

Switches off the device function.

START_X       X=1;2;4;5;7

                    X=80; 90

Switching on of device - (remote) function

Activation of the previous watchdog function. If no target value is entered within 

60 seconds, the (old function) heating is switched off.

Remote

STOP_X        X=1;2;4;5;7 

                    X=80; 90

Switching off of device - (remote) function The variables set with OUT_SP_X remain 

saved. Includes command RMP_STOP.

Remote

background image

32

“Watchdog” functions; monitoring of the serial data flow

If, once this function has been activated (see NAMUR commands), 

there is no retransmission of the command from the computer 

within the set time (“watchdog time”), the heating and stirring 

functions are switched off in accordance with the set “watchdog” 

function or are changed to the set target values.

The data transmission may be interrupted by, for example, a crash 

in the operating system, a power failure in the PC or an issue with 

the connection table between the computer and REC control.

“Watchdog” - mode 1

If there is an interruption in data communications (longer than 

the set watchdog time), the heating and stirring functions are 

switched off and ER 2 is displayed.

“Watchdog” - mode 2

If there is an interruption in data communications (longer than 

the set watchdog time), the speed target value is changed to 

the WD safety speed limit and the temperature target value is 

changed to the WD safety temperature limit value. The warning 

WD is displayed.

Connection possibilities between the RET control 

and external devices

The following adapters and adapter cables are available from 

IKA

®

for connecting the RET control to external devices.

PC 1.1 cable

The PC 1.1 cable is used to connect the 9-pin plug to a computer.

 Accessories

 Troubleshooting

The device will indicate that there is a fault by displaying an error 

message “Er X” (see table) on the screen.

Heating will be switched off permanently (exception: WD - see 

table).

The stirring function continues to operate at the pre-set speed if 

there is a fault in the heating circuit (exception: WD - see table).

In order to reset the device after a fault, it may need to be discon-

nected from the power supply for several seconds. The period the 

device is switched off may also need to be longer if the tempera-

tures need to be counterbalanced (residual heat display may no 

longer be active; see “Residual heat display”).

The safety circuit test takes place within 20 seconds of switching 

on the device; after the test, the other monitoring functions are 

activated.

If a fault cannot be eliminated directly, you must perform a 

RESET 

 (see “Settings” menu)

. If the faults still cannot be elimi-

nated, the appliance must be inspected by a technical service 

technician.

• Magnetic rods:

  

 ø   6 mm; max. length 15 mm

  

 ø   7 mm; max. length 60 mm

  

 ø 10 mm; max. length 80 mm

• 

RS 1

  

Set of magnetic stirring bars

• 

RSE

  

Stirring bar remover

• 

H 1000

  

Beaker 

1 l

• 

H 1500

  

Beaker 

1.5 l

• 

H 3000

  

Beaker 

3 l

• 

H 5000

  

Beaker 

5 l

• 

H 8000

  

Beaker 

8 l

• 

H 135.3

  

Flask carrier

• 

H 135.310

  

Quarter 20 ml reaction vessel

• 

H 135.311

  

Quarter 30 ml reaction vessel

• 

H 135.312

  

Quarter 40 ml reaction vessel

• 

H 135.313

  

Quarter 4 ml reaction vessel

• 

H 135.314

  

Quarter 8 ml reaction vessel

• 

H 135.315

  

Quarter 16 ml reaction vessel

• 

H 135.4

  

Reaction block for 100 ml round flask

• 

H 135.410

  

Insert for 10 ml round flask

• 

H 135.411

  

Insert for 25 ml round flask

• 

H 135.412

  

Insert for 50 ml round flask

• 

H 135.5

  

Reaction block for 500 ml round flask

• 

H 135.510

  

Insert for 200 ml round flask

• 

H 135.511

  

Insert for 250 ml round flask

• 

H 135.512

  

Insert for 300 ml round flask

• 

H 135.6

  

Reaction block for 1,000 ml round flask

• 

H 135.610

  

Insert for 500 ml round flask

• 

H 16 V

  

Support rod

• 

H 16.1

  

Extension

• 

H 38

  

Holding rod

• 

H 44

  

Boss head clamp

• 

PT100.50

   

Temperature sensor

• 

PT100.51

   

Temperature sensor, glass coated 

• 

PT1000.50

   

Double temperature sensor

• 

PT1000.51

   

Double temperature sensor, glass coated

  

for further accessories see 

www.ika.com

.

1

RxD 2

TxD 3

4

GND 5

6

RTS 7

CTS 8

9

PC

1

2

3

4

5

6

7

8

9

1

2 RxD

3 TxD

4

5 GND

6

7 RTS

8 CTS

9

5

4

3

2

1

9

8

7

6

background image

33

Error messages and troubleshooting table

Fault

Description

Causes

Effect

Solutions

Er 2

In remote operation (PC) mode 

with activated watchdog func-

tion in mode 1: no commu-

nication between PC and RET 

control

•   PC does not transmit any data 

within the set watchdog time

•   Connection to PC interrupted

Heating switched off

Motor switched off

•   Change watchdog time

•   Transmit data from PC within set 

watchdog time (OUT_WDx@m)

•   Check cable and plug

Er 3

Device internal temperature 

exceeds 80°C

•   Heat accumulation between 

heating plate and housing

•   Permitted ambient temperature 

exceeded

Heating switched off

•   Switch device off, leave to cool 

and switch on again

•   Change experiment

•   Observe maximum permissible 

ambient temperature

Er 4

Difference between target and 

actual speed exceeds 300 rpm

•   Motor blocked or overloaded

Heating switched off

Motor switched off

•   Reduce load torque or use smaller 

magnetic rods

•   Reduce target speed

Er 5

No temperature increase in 

sensor at continuous tempera-

ture difference (after approx. 

3 mins)

•   Sensor not in medium

•   Volume of medium to be mea

-

sured too large

•   Heat conductivity of medium to 

be measured too low

•   Heat conductivity of the vessel is 

too low

•   In the case of indirect heating, 

the overall heat conductivity 

resistance is too large

Heating switched off

•   Place the sensor in the medium

•   Reduce the volume of the media

•   Use a carrier fluid with better heat 

conductivity properties

•   Replace the glass vessel with a 

metal pot

Er 6

Interruption in safety circuit

•   Defective connection cable

Heating switched off

•   Replace cable

Er 13

SAFE TEMP P adjustable safety 

circuit defective

•   Target/current different of the 

adjustable safety circuit for mini-

mum temperature monitoring

Heating switched off

•   After switching on, change the 

SAFE TEMP to a different value; if 

this solves the issue, the previous 

value can be reset by switching 

the device off and on again

Er 14

Short circuit in temperature 

sensor 1

•   Short circuit in temperature sen

-

sor plug

•   Short circuit in the cable or tem

-

perature sensor

Heating switched off

•   Check the plug

•   Replace the temperature sensor

Er 21

Fault during heating plate sa-

fety test

•   Safety relay does not open

Heating switched off

•   Switch device off, leave to cool 

and switch on again

Er 22

Fault during heating plate sa-

fety test

•   S_CHECK cannot generate 

H_S_TEMP

Heating switched off

•   Switch device off, leave to cool 

and switch on again

Er 24

The temperature of the heating 

plate exceeds SAFE TEMP H 

(Hotplate)

•   SAFE TEMP H (Probe) has been 

set to lower than the current 

temperature of the heating plate

•   Disconnection of heating plate 

control temperature sensor

Heating switched off

•   Leave the heating plate to cool

•   Set the SAFE TEMP H (Hotplate) 

higher

Er 25

Heating switching element 

monitoring

•   Heater control circuit switch 

(TRIAC) short-circuited

•   Safety relay has interrupted the 

heating circuit

•   Heater or the supply line is dis

-

connected

•   Interruption in heating plate 

safety temperature sensor

Heating switched off

•   Switch device off, leave to cool 

and switch on again

Er 26

Difference between internal 

heating plate control and safe-

ty temperature sensor too large

•   Irregular temperature distribu

-

tion across heating plate due to 

sporadic heat dissipation

•   Defective control or safety tem

-

perature sensor

Heating switched off

•   Switch device off, leave to cool 

and switch on again

•   Ensure regular heat dissipation 

when using metal blocks, etc. 

on the flat surface of the heating 

plate

Er 31

Defective monitoring of the 

heating control circuit (TRIAC)

Heating switched off

•   Contact customer service

Er 41

Bar breakaway

•   The bar has broken away four 

times in a row without an explicit 

change in the target value

Target speed reduced 

by at least 300 rpm

Heating switched off

•   Select a lower target speed

•   Select a smaller stirring bar

background image

34

 Maintenance and cleaning

The device should operate without the need for maintenance, 

however it is subject to natural wear and tear on parts and their 

statistical failure rate.

Cleaning

Remove the device from the mains 

before cleaning.

Only use cleaning materials recommended by 

IKA

®

.

Dirt

Cleaning agent

Dyes

Isopropyl alcohol

Building materials

  Water containing detergent/isopropyl alcohol

Cosmetics

Water containing detergent/isopropyl alcohol

Food

Water containing detergent

Fuels

Water containing detergent

- Ensure no liquid enters the device during cleaning.

-  Wear protective gloves when cleaning the device.

-  Please consult with 

IKA

®

 before using any cleaning or decon-

tamination methods not specifically recommended.

Ordering spare parts

When ordering spare parts, please make sure to indicate the fol-

lowing:

-  device type

-  device manufacturing number; see rating plate

-  software version (second value displayed when switching on 

the device)

-  position number and description of spare part; see 

www.ika.

com

.

Repairs

Please only send devices in for repair that have been 

cleaned and are free of materials which might present 

health hazards.

For this, use the “

certificate of compliance

” form which you can 

obtain from 

IKA

®

 or can download a version for printing from the 

IKA

®

 website at 

www.ika.com

.

If your appliance requires repair, return it in its original packaging. 

Storage packaging is not sufficient when sending the device - also 

use appropriate transport packaging.

background image

35

 Technical data

Rated voltage 

V AC 

220 - 230±10%

   or 

V AC 

115±10%

   or 

V AC 

100±10%

Frequency 

Hz 

50/60

Display 

TFT

Timer 

99 h 59 min 59 sec

Stirring

Motor: 

ball bearing mounted, brushless EC motor

Max. motor power consumption 

22

Max. motor output 

12

Speed range 

rpm 

0; 50 to 1,700

Target speed setting resolution 

rpm 

1

Actual speed display resolution 

rpm 

1

Speed variation 

2

  (without load, 115/230 V, at 1,700 rpm, RT 25 °C)

Stirring capacity (H

2

O) 

20

Heating

Heating plate material 

Stainless steel 1.4301

Heating plate diameter 

mm 

135

Heat output (100/115/230 V) 

600±5%

Temperature range 

°C 

RT - 340

Target temperature setting resolution 

0.1

Actual temperature display resolution

- of the medium temperature 

0.1

- of the carrier fluid temperature 

1

Adjustable heating plate safety circuit 

°C 

50 - 380 (±5°C)

Safety circuit setting resolution 

Temperature sensor in medium 

PT 100 or 2xPT 1000, DIN EN 60751 Class A

Temperature sensor deviation 

<=±

[

0.15 + 0.002°(T)

]

  (PT 100 or 2xPT 1000, DIN EN 60751 Class A)

Deviation in section of temperature measurements 

<=±0.3 < 100°C

    

<=±0.5 > 100°C

Deviation in section of temperature measurements + sensor 

<=±0.45 at 0°C

    

<=±0.85 at 100°C

Deviation in heating plate temperature 

±3 without vessel; heating plate at 100°C

Control hysteresis with temperature sensor 

±0.2

  (500 ml water, 600 ml glass beaker,

  30 mm bar, 800 rpm, PT 100.50, 50°C)

Heating plate control hysteresis 

±3 without vessel; heating plate at 100°C

Permissible ambient temperature 

°C 

+5 to +40

Permissible relative humidity 

80

Permissible operating time 

100

Protection class according to DIN EN 60529 

IP 42

Weighing range 

0 - 5,000

Weighing range tolerance 

< 500: ±1

    

> 500: ±5 

Level of contamination 

2

Overvoltage category 

II

Protection class 

1 (protective earth)

Use above sea level 

max. 2,000

Dimensions (W x D x H) 

mm 

160 x 270 x 85

Weight 

kg 

2.7

             

Subject to technical changes

background image

36

FR

Table des matières

  

  Page

Éléments de commande et écran 

    2

Table des matières 

   36

Déclaration de conformité CE 

   36

Garantie 

   36

Explication des symboles 

   36

Consignes de sécurité 

   37

Déballage 

   38

Utilisation conforme 

   38

Montage de la tige du trépied 

   38

Réglage de la limite de la température de sécurité  

 38

 Explication des symboles

DANGER

PRUDENCE

AVERTISSEMENT

Le présent symbole signale des informations 

cruciales pour votre sécurité et votre santé

. Le non-res-

pect de ces indications peut nuire à la santé et causer des blessures.

Le présent symbole signale des informations importantes 

pour le bon fonctionnement technique de 

l’appareil

.

Le non-respect de ces indications peut endommager l’appareil.

Le présent symbole signale des informations 

importantes pour le bon fonctionnement de l’appa-

reil et pour sa manipulation

.

Le non-respect peut avoir pour conséquence des résultats de mesure 

imprécis.

ATTENTION - remarque sur une mise en danger en raison du magnétisme.

DANGER - Remarque sur une mise en danger en raison d’une surface chaude.

Avertissement de danger générique.

ATTENTION

DANGER

 Déclaration de conformité CE

Nous déclarons sous notre seule responsabilité que ce produit est conforme aux réglementations des directives 2006/95/CE et  

2004/108/CE et est en parfait accord avec les normes et documents normatifs suivants :

DIN EN IEC 61010-1, -2-010, -2-051 et DIN EN IEC 61326-1.

 Garantie

Conformément aux conditions de vente et de livraison d’

IKA

®

, la garantie a une durée de 24 mois. En cas de demande de garantie, 

s’adresser au distributeur ou expédier l’appareil accompagné de la facture et de la raison de la réclamation directement à notre usine. 

Les frais de port sont à la charge du client.

La garantie ne s’étend pas aux pièces d’usure et n’est pas valable pour les erreurs causées par une manipulation non conforme, un en-

tretien et une maintenance insuffisants ou le non-respect des instructions du présent mode d’emploi.

Utilisation 

  39

Réglage des modes d’exploitation 

  39

Fonctions 

  40

Ports et interfaces 

  45

Accessoires 

  48

Défauts 

  48

Entretien et nettoyage 

  50

Caractéristiques techniques 

  51

background image

37

>100 mm

>100 mm

>100 mm

Consignes de sécurité

•  

Lire entièrement le mode d’emploi avant la mise en service 

et respecter les consignes de sécurité.

•  Conserver le mode d’emploi de manière à ce qu’il soit accessible à 

tous.

Indications générales

•  

Veiller à ce que seul un personnel formé travaille avec l’appareil.

•  Respecter  les  consignes  de  sécurité,  les  directives,  ainsi  que  les 

mesures de prévention des accidents.

•  

La prise doit être mise à la terre (contact à conducteur de protection).

  

Attention - magnétisme !

  

Tenir compte des effets du champ magné-

tique (pacemaker, support de données...).

  

Risque de brûlure !

   Toucher prudemment les pièces du loge 

  ment et la plaque chauffante. La plaque 

chauffante peut atteindre une chaleur supérieure à 320 °C. Atten-

tion à la chaleur résiduelle après la mise hors tension!

   

L’appareil doit être transporté à froid !

Installation de l’appareil

  

N’utilisez pas

 l’appareil dans des atmos-

phères explosives, avec des matières dan-

gereuses et sous l’eau.

•  Placez l’appareil à un endroit dégagé sur une surface plane, stable, 

propre, non glissante, sèche et non inflammable.

•  Les pieds de l’appareil doivent être propres et ne pas être abîmés.

•  Le câble secteur et les câbles des sondes externes ne doivent pas 

entrer en contact avec la plaque d’appui chauffante.

•  Avant chaque utilisation, contrôlez l’état de l’appareil et des 

accessoires. N’utilisez pas de pièces endommagées.

Milieux autorisés/Impuretés/ Réactions

   

  

Attention!

 Avec cet appareil, ne peu-

vent être traités et chauffés que des 

agents dont le point d’inflammation dé-

passe la température de sécurité limite 

(50 ... 380 °C).

La limite de température de sécurité fixée doit toujours être au 

moins de 25 °C inférieure au point d’inflammation de l’agent utilisé.

   

  

Attention aux risques suivants :

   -  Des matériaux inflammables

  -  Des milieux combustibles à faible pression de vapeur

  -  bris de verre

  -  dimensionnement du récipient erroné

  -  niveau de remplissage du milieu trop élevé

  -  état du récipient non stable.

•  Ne traitez des substances pathogènes que dans des récipients fer

-

més et sous une hotte d’aspiration adaptée. En cas de questions, 

contactez 

IKA

®

.

Ne traitez que des substances pour lesquel-

les l’apport d’énergie pendant l’opération 

ne pose pas problème. Ceci s’applique 

également aux autres apports d’énergie, par ex. la radiation lu-

mineuse.

•  En cas de vitesses de rotation élevées, la plaque de composants 

peut  s’échauffer  même  sans  chauffage  par  l’aimant  d’entraîne

-

ment. 

•  Attention  aux  salissures  et  réactions  chimiques  éventuelles  non 

souhaitées.

• 

 Éventuellement, des particules d’abrasion provenant de l’appareil ou 

des pièces en rotation puissent se retrouver dans le milieu à traiter.

•  En cas d’utilisation de barreaux aimantés revêtu, tenir compte de 

ce qui suit  : 

Les réactions chimiques du PTFE se produisent au 

contact avec des métaux alcalins et alcalino-terreux fondus ou 

dissous ainsi qu’avec des poudres fines des métaux du 2ème et 

3ème groupe du système périodique à des températures supé-

rieures à 300-400 °C. Seul le fluor élémentaire, le trifluorure de 

ATTENTION

DANGER

chlore et les métaux alcalins l’affectent; les hydrocarbures halo-

gènes ont un effet de gonflement réversible

.

  

  

Source : Römpps Chemie-Lexikon et « Ullmann » Vers. 

19)

•  En cas d’utilisation des métaux alcalins et alcalino-terreux dissous 

et à des températures supérieures à 250 °C, il est essentielle de 

n’utiliser que de barreaux aimantés !

Réalisations des essais

  

Portez

 l’équipement de protection per-

sonnel en fonction de la classe de danger 

de la substance à traiter. Sinon, il y a dan-

ger de :

  -  Pulvérisation et évaporation de liquides 

  -  éjection de pièces

  -  Libération de gaz toxiques ou inflammables.

•  Réduisez la vitesse de rotation si :

  -  la substance est projetée hors du récipient, parce que la vitesse 

de rotation est trop élevée

  -  le fonctionnement est irrégulier

  -  le récipient se déplace sur la plaque d’appui.

Accessoires

•  Un travail en toute sécurité n’est garanti qu’avec les accessoires 

décrits dans le chapitre « 

Accessoires

 ».

•  Ne montez d’accessoires que lorsque la fiche secteur est débran

-

chée. 

•  Respectez le mode d’emploi des accessoires.

•  Immerger la sonde de température externe (PT1000, ETS-D ....) 

dans le milieu à une profondeur d’au moins 20 mm.

•  La sonde de température externe PT1000 branchée doit toujours 

se trouver dans le milieu.

•  Les accessoires doivent être correctement connectés de façon sûre 

à l’appareil et ne doivent pas se détacher seuls. Le barycentre de la 

structure doit se trouver dans la surface d’appui.

Alimentation en tension/coupure de l’appareil

•  L’indication de la tension de la plaque signalétique doit coïncider 

avec la tension du réseau.

•  La prise de courant utilisée pour le branchement sur secteur doit 

être facile d’accès.

•  Il n’est possible de couper l’alimentation en courant de l’appareil 

qu’en débranchant la prise secteur ou de l’appareil.

  

Après une interruption de l’alimentation 

électrique, l’appareil redémarre automa-

tiquement en mode B.

Pour protéger l’appareil

•  L’appareil ne doit être ouvert que par un spécialiste.

•  Ne couvrez pas l’appareil, même pas partiellement, par exemple 

avec des plaques ou des films métalliques. Le résultat est une sur-

chauffe.

•  Évitez les chocs et les coups sur l’appareil ou sur les accessoires.

•  Veillez à obtenir une plaque d’appui propre !

•  Respecter les distances minimales entre les appareils, entre l’appa

-

reil et le mur et au-dessus de la structure (min. 800 mm).

AVERTISSEMENT

DANGER

AVERTISSEMENT

AVERTISSEMENT

AVERTISSEMENT

DANGER

background image

38

Déballage

• Déballage

  -  Déballez l’appareil avec précaution

  -  En cas de dommages, établissez immédiatement un constat 

correspondant (poste, chemin de fer ou transporteur)

• Étendue de la livraison

-  Agitateur magnétique réchauffable RET control visc

  -  Câble secteur

  -  Mode d’emploi

  -  Tournevis (circuit de sécurité)

  -  Câble USB

  -  Capot de protection 

H104

  -  Barreau aimanté 

IKAFLON

30

 et 

40

mm

  -  Sonde de température 

PT 100.70

 Utilisation conforme

 Montage de la tige de trépied

Remarque :

Pour travailler avec les cuves de plus de 180 mm Ø, 

veuillez utiliser la tige 

de trépied H de 16  V 

avec la console H 16.1.

La limite de la température de sécurité définie s’affiche en 

permanence

 (7

).

Réglage par défaut : env. 380 °C

Plage de réglage : 50 - 380 °C

Réglage de la limite de la température de sécurité

a

   

Portez l’interrupteur de l’appareil 

(A)

 en position ON

a

   

Régler la limite de la température de sécurité au point 

(B)  

en utilisant le tournevis fourni

Affichage de la valeur réglée à l’écran

La température maximale de la plaque chauffante de 340 °C Max est 

au moins 10 °C inférieure à la limite de température de sécurité réglée. 

Attention  : La limite de température de sécurité réglée 

doit toujours être au moins de 25  °C inférieure au point 

d’inflammation de l’agent à traiter.

Test de fonctionnalité du disjoncteur de sécurité

a

   

Chauffez l’appareil à plus de 50 °C.

a

   

Placez la limite de la température de sécurité en butée gauche 

(50 °C) Portez l’interrupteur de l’appareil (A) en position ON

Affichage sur l’écran        Err 25

 Réglage de la limite de la température de sécurité

Ne tournez pas la vis au-delà des butées gauche et droite sous 

peine de détruire le potentiomètre.

• Utilisation

   -  Pour mélanger et / ou chauffer des liquides.

• Zone d’utilisation (uniquement en intérieur)

   -  Laboratoires  

- Écoles

   -  Pharmacies  

- Universités

L’appareil est adapté à une utilisation dans toutes les zones, ex-

ceptées les zones résidentielles qui sont directement raccordées à 

un réseau d’alimentation basse tension alimentant des zones rési-

dentielles.

La protection des l’utilisateur n’est plus assurée si l’appareil est uti-

lisé avec des accessoires non fournis ou non recommandés par le 

fabricant, ou si l’appareil est utilisé de manière non conforme, en 

ne respectant pas les prescriptions du fabricant, ou si des modifi-

cations ont été effectuées sur l’appareil ou le circuit imprimé par 

un tiers.

•  

Enlever le bouchon fileté

 (P)

•  Détachez le capuchon de protection de la tige du 

trépied

•  Mettez la rondelle entre le boîtier et l’écrou

•  Vissez la tige du trépied à la main jusqu’à la butée

•  Vissez l’écrou avec une clé à fourche de 17

•  Monter les accessoires avec de manchons en croix

background image

39

Retirez le film protecteur de la plaque d’appui ! Le changement dans le menu est effectué à l’aide de la touche de menu « Back ».

Mise en 

service

Agiter

Chauffer

Raccordement 

du thermomètre 

externe

(régulation directe 

de la température 

dans le milieu)

Peser

Mesure du pH

Minuteur

Amener l’interrupteur de l’appareil (A) en position OFF

Brancher le câble secteur dans la prise secteur (K)

Amener l’interrupteur de l’appareil (A) en position ON

Le mode standard A est défini (voir « modes d’exploitation »)

Régler la vitesse de rotation du moteur avec le bouton de commande (D)

La valeur réglée s’affiche à l’écran (F)

Démarrer la fonction d’agitation en appuyant sur le bouton de commande (D)

i

Le symbole (16) s’affiche quand le moteur est activé

  F 

Régler la limite de la température de sécurité (voir « Réglage de la limite de la température de sécurité »)

  F 

Régler la température de consigne avec le bouton de commande (C)

  a 

La valeur réglée s’affiche à l’écran (J)

  F 

Régler le mode de régulation de température (voir « Réglage du mode de régulation de température »)

  F 

Démarrer la fonction de chauffage en appuyant sur le bouton de commande (C)

    

La température de consigne et la température réelle s’affichent en permanence à l’écran (J) :

i

Lorsque le chauffage est allumé, le symbole de chauffage (4) s’affiche

    

Le température max. de la plaque chauffante est de 340 °C

i

Tant que la température de la surface d’appui dépasse 50 °C,  

  et la température courante de la plaque chauf-

fante s’affichent sur l’écran (J), même lorsque l’appareil est éteint

Amener l’interrupteur de l’appareil (A) en position OFF

Connecter le thermocontact de sécurité selon DIN 12878 classe 2 ou la sonde de température PT 100 (sonde indivi-

duelle et/ou sonde double PT1000) à la douille (L)

Amener l’interrupteur de l’appareil (A) en position ON

i

   Sonde de température PT 100 

a  

La température réelle (5) affichée sur l’écran (J) correspond à la température 

du milieu. Le symbole « échantillon » (11) s’allume.

i  

Sonde de température PT 1000 

a  

La température réelle (5) de la sonde de température affichée sur l’écran (J) cor-

respond à la température du milieu. Le symbole « sonde de température externe » 

(11) s’allume.

    

    

 Dans le même temps, la température du liquide porteur (3) s’affiche. 

Appuyer sur la touche (G)

attendre que le système soit stable (une fois que le système est stable, un « g » s’affiche derrière le chiffre 0.

Placer le poids au milieu de la plaque chauffante et attendre que le système soit stable.

a

La plage d’affichage maximale est de 2000 g 

a

La plage de pesage maximale est de 5000 g 

a

La fonction Peser peut être activée uniquement quand les fonctions Chauffer et Agiter sont désactivées

a

Le module de pesage peut être calibré (voir le chapitre « Étalonnage du module de pesage »)

a

Voir le chapitre « Étalonnage de la sonde pH et mesure du pH »

Appuyer sur la touche (H) 

Régler le temps désiré en heures, minutes et secondes en tournant le bouton rotatif (D) et confirmer en appuyant.  

Après le démarrage des fonction Chauffer ou Agiter, la fonction minuterie commence en comptant en arrière (voir 

« fonction minuterie »)

i  

Lorsque la minuterie atteint 0, la fonction Chauffer est toujours désactivée, même si le moteur a été démarré en premier.

    

Si le chauffage est activé pendant le fonctionnement de la minuterie, seulement le chauffage est désactivé, le 

moteur continue de tourner (retard d’ébullition). 

Utilisation 

(notice)

 Réglage des modes d’exploitation

Exploitation de l’appareil avec les modes A, B ou C

Réglage d’usine : Mode A

Le mode réglé s’affiche à l’écran (10) en permanence.

Mode A

Tous les valeurs définies sont conservées après l’arrêt ou le 

débranchement de l’appareil. Après la mise en marche de l’appareil, 

l’état des fonctions Chauffer et Agiter est désactivé (OFF).

Mode B

Tous les valeurs définies sont conservées après l’arrêt ou le 

débranchement de l’appareil. Après la mise en marche de 

l’appareil, l’état des fonctions Chauffer et Agiter avant le dernier 

arrêt est reprise (ON ou OFF).

background image

40

Mode C

Ce mode permet de travailler avec ou sans sonde de 

température.

Après la mise en marche de l’appareil avec l’interrupteur de 

l’appareil, les fonctions Agiter et Chauffer démarrent avec le 

dernier état et les dernières valeurs. Les valeurs de consigne 

réglées pour les modes de fonctionnement A et B ne peuvent 

pas être modifiées.

Changement des modes d’exploitation

-  

Faites défiler le menu jusqu’à l’icône « Mode »

-  

Ouvrez le sous-menu en appuyant sur le bouton rotatif (D)

-  

Sélectionnez le mode en tournant le bouton rotatif (D) et con-

firmez le mode sélectionné en appuyant sur le bouton rotatif 

(D)

-  

Une brève description du mode sélectionné s’affiche

-  

Confirmez le mode en appuyant sur le bouton rotatif (D).

 Fonctions

Le mode de régulation de la température sélectionné pendant le 

fonctionnement avec PT100 est signalé par le symbole 

 (11) 

dans l’affichage digital.

La température de consigne (2) affichée 

et la température réelle (5) se réfèrent 

à la température du milieu. La tempé-

rature de la plaque chauffante est limi-

tée par la température de sécurité (7) 

réglée. La régulation de la température 

du milieu s’effectue via un microcon-

trôleur. Cela permet de chauffer la tem-

pérature du milieu détectée par la PT 

100 le plus rapidement possible et sans 

dépassement à la température réglée. 

Le microcontrôleur s’adapte automa-

tiquement aux différentes capacités 

thermiques des différentes récipients 

à poser utilisés et à leur contenu. Cela 

garantit toujours une régulation opti-

male de la température sans dérive en 

température et ondulation.

Réglage de la température du milieu avec la 

sonde de température PT100

L’utilisateur peut personnaliser la régulation de température selon 

ses besoins dans trois modes différents. Le mode de régulation de 

température sélectionné s’affiche en plus du symbole (

11).

Modes de régulation

Changement des modes de 

régulation

-  

Faites défiler le menu 

jusqu’à l’icône « Chauffer »

-  

Ouvrez le sous-menu Mé-

thode de régulation en 

appuyant sur le bouton 

rotatif (D)

-  

Sélectionnez le mode en 

tournant le bouton rota-

tif (D) et confirmez le mode 

sélectionné en appuyant 

sur le bouton rotatif (D)

-  

Une brève description du 

mode sélectionné s’affiche

-  

Confirmez le mode en 

appuyant sur le bouton 

rotatif (D).

PID

:  pour les milieux très sensibles qui doivent être traités avec 

prudence.

  Avantage :

  •    chauffage très précis du 

liquide

  •    aucun dépassement de 

la température

  Désavantage :

  •    fonctionnement de 

chauffage le plus lente

T

Set

t

Fast PID:  

Combine les avantages de PID et de la régulation 

à 2 points 

(chauffage très 

rapide et de 

haute précision 

de régulation, les 

dépassements 

sont possibles). 

Régulation à 2 points

 :

     Pour les milieux qui ne sont pas trop sensibles.

  Avantage :

  •    Le chauffage est beau

-

coup plus rapide que le 

PID

  Désavantage :

  •    Dépassement de la 

température du milieu 

jusqu’à 10 °K

T

Set

t

T

Set

t