Labconco VACUUBRAND Diaphragm Pump 7393001: Troubleshooting

Troubleshooting: Labconco VACUUBRAND Diaphragm Pump 7393001

background image

page 50 of 74

Fault

Possible cause

Remedy

❑ 

Pump does not 

start or stops im-

mediately.

➨

Electrical power cord 

not plugged in, electri-

cal supply failure?

✔

Plug in power cord. 

Check fuse.

➨

Device fuse blown?

✔

Identify cause of failure. 

Replace device fuse.

➨

Overpressure in outlet 

line?

✔

Remove blockage in line, 

open valve.

➨

Motor overloaded?

✔

Allow motor to cool down, 

identify and eliminate 

cause of failure. Manual 

reset is necessary. Switch 

off pump or unplug.

❑

Pump does not 

achieve its ultimate 

vacuum or usual 

pumping speed.

➨

Centring ring at small 

flange connection not 

correctly positioned, or 

leak in the pipeline or 

vacuum system?

✔

Check pump directly - 

connect vacuum gauge 

directly at pump inlet - 

then check connection, 

pipeline and vacuum 

system if necessary.

➨

Long, narrow vacuum 

line?

✔

Use lines with larger di-

ameter, length as short as 

possible.

➨

Pump has been ex-

posed to condensate?

✔

Allow pump to run for 

some minutes with atmo-

spheric pressure at the 

inlet to purge.

➨

Deposits have been 

formed inside the 

pump?

✔

Clean and inspect the 

pump heads.

➨

Diaphragms or valves 

damaged?

✔

Replace diaphragms and/

or valves.

➨

Outgassing substances 

or vapor generated in 

the process?

✔

Check process parame-

ters.

Troubleshooting

background image

 page 51 of 74

Fault

Possible cause

Remedy

❑

Pump too noisy.

➨

Atmospheric or high 

pressure at the pump 

inlet?

✔

Connect hose or silencer 

to pump outlet. Be careful 

not to cause outlet over-

pressure, especially with 

condensable vapors.

➨

Diaphragm crack or 

diaphragm clamping 

disc loose?

✔

Perform maintenance.

➨

Other than above men-

tioned causes?

✔

Contact local distributor.

❑

Pump seized.

✔

Contact local distributor.

➨

 A service manual with exploded view drawings, spare parts list and 

directions for repair is available on request.

+ 

The service manual is intended for trained service people only.